Directive 2007/47/EC of the European Parliament and of the Council of 5 September 2007 amending Council Directive 90/385/EEC on the approximation of the laws of the Member States relating to active implantable medical devices, Council Directive 93/42/EEC concerning medical devices and Directive 98/8/EC concerning the placing of biocidal products on the market (Text with EEA relevance)
Directive 2007/47/EC of the European Parliament and of the Council of 5 September 2007 amending Council Directive 90/385/EEC on the approximation of the laws of the Member States relating to active implantable medical devices, Council Directive 93/42/EEC concerning medical devices and Directive 98/8/EC concerning the placing of biocidal products on the market (Text with EEA relevance)
Directive 2007/47/EC of the European Parliament and of the Council of 5 September 2007 amending Council Directive 90/385/EEC on the approximation of the laws of the Member States relating to active implantable medical devices, Council Directive 93/42/EEC concerning medical devices and Directive 98/8/EC concerning the placing of biocidal products on the market (Text with EEA relevance)
IO L 247, 21.9.2007, p. 21–55
(BG, ES, CS, DA, DE, ET, EL, EN, FR, IT, LV, LT, HU, MT, NL, PL, PT, RO, SK, SL, FI, SV) Foilsíodh an doiciméad seo in eagrán speisialta
(HR)
No longer in force, Date of end of validity: 25/05/2021; Arna aisghairm go hintuigthe ag 32017R0745 Féach 32020R0561
21.9.2007 |
EN |
Official Journal of the European Union |
L 247/21 |
DIRECTIVE 2007/47/EC OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL
of 5 September 2007
amending Council Directive 90/385/EEC on the approximation of the laws of the Member States relating to active implantable medical devices, Council Directive 93/42/EEC concerning medical devices and Directive 98/8/EC concerning the placing of biocidal products on the market
(Text with EEA relevance)
THE EUROPEAN PARLIAMENT AND THE COUNCIL OF THE EUROPEAN UNION,
Having regard to the Treaty establishing the European Community, and in particular Article 95 thereof,
Having regard to the proposal from the Commission,
Having regard to the opinion of the European Economic and Social Committee (1),
Acting in accordance with the procedure laid down in Article 251 of the Treaty (2),
Whereas:
(1) |
Council Directive 93/42/EEC (3) requires the Commission to submit a report to the Council, no later than five years from the date of implementation of that Directive, concerning: (i) information on incidents occurring following the placing of devices on the market, (ii) clinical investigation carried out in accordance with the procedure set out in Annex VIII to Directive 93/42/EEC, and (iii) design examination and EC type examination of medical devices that incorporate, as an integral part, a substance which, if used separately, may be considered to be a medicinal product as defined in Directive 2001/83/EC of the European Parliament and of the Council of 6 November 2001 on the Community code relating to medicinal products for human use (4) and which is liable to act upon the body with action ancillary to that of the device. |
(2) |
The Commission brought forward the conclusions of that report in its Communication to the Council and the European Parliament on medical devices which, at the request of the Member States, was expanded to cover all aspects of the Community regulatory framework for medical devices. |
(3) |
This Communication was welcomed by the Council in its Conclusions on medical devices of 2 December 2003 (5). It was also discussed by the European Parliament which on 3 June 2003 adopted a resolution on the health implications of Directive 93/42/EEC (6). |
(4) |
Following from the conclusions drawn in that Communication it is necessary and appropriate to amend Council Directive 90/385/EEC (7), Directive 93/42/EEC and Directive 98/8/EC of the European Parliament and of the Council (8). |
(5) |
To ensure consistency of interpretation and implementation between Directives 93/42/EEC and 90/385/EEC the legal framework related to issues such as authorised representative, the European databank, health protection measures, and the application of Directive 93/42/EEC as regards medical devices incorporating stable derivates of human blood or human plasma, as introduced by Directive 2000/70/EC (9), should be extended to Directive 90/385/EEC. The application of the provisions on medical devices incorporating stable derivates of human blood or human plasma includes application of Directive 2002/98/EC of the European Parliament and of the Council of 27 January 2003 setting standards of quality and safety for the collection, testing, processing, storage and distribution of human blood and blood components and amending Directive 2001/83/EC (10). |
(6) |
It is necessary to clarify that software in its own right, when specifically intended by the manufacturer to be used for one or more of the medical purposes set out in the definition of a medical device, is a medical device. Software for general purposes when used in a healthcare setting is not a medical device. |
(7) |
Particular care should be taken to ensure that the reprocessing of medical devices does not endanger patients’ safety or health. It is therefore necessary to provide clarification on the definition of the term ‘single use’, as well as to make provision for uniform labelling and instructions for use. Moreover, the Commission should engage in further analysis in order to see if additional measures are appropriate to ensure a high level of protection for patients. |
(8) |
In the light of technical innovation and the development of initiatives at the international level it is necessary to enhance the provisions on clinical evaluation, including clarification that clinical data is generally required for all devices regardless of classification and the possibility to centralise data on clinical investigations in the European databank. |
(9) |
In order to provide clearer evidence of the compliance of custom-made device manufacturers, an explicit requirement for a post market production review system involving incident reporting to authorities should be introduced, as is already in place for other devices, and to enhance patient information, a requirement should be introduced that the ‘Statement’ under Annex VIII to Directive 93/42/EEC should be available to the patient and that it should contain the name of the manufacturer. |
(10) |
In the light of technical progress in information technology and medical devices, a process should be provided to allow information supplied by the manufacturer to be available by other means. |
(11) |
Manufacturers of Class I sterile and/or measuring medical devices should be given the option of using the full quality assurance conformity assessment module in order to provide them with more flexibility in the choice of compliance modules. |
(12) |
In order to support market surveillance activities by Member States it is necessary and appropriate, in the case of implantable devices, to increase the time period for the retention of documents for administrative purposes to at least 15 years. |
(13) |
For the appropriate and efficient functioning of Directive 93/42/EEC as regards regulatory advice on classification issues arising at national level, in particular on whether or not a product falls under the definition of a medical device, it is in the interest of national market surveillance and the health and safety of humans to establish a procedure for decisions on whether or not a product falls under the medical device definition. |
(14) |
To ensure that, where a manufacturer does not have a registered place of business in the Community, authorities have a single individual person authorised by the manufacturer whom they can address in matters relating to the compliance of the devices with the Directives it is necessary to introduce an obligation for such manufacturers to designate an authorised representative for a device. This designation should be effective at least for all devices of the same model. |
(15) |
To further ensure public health and safety it is necessary to provide for a more consistent application of the provisions on health protection measures. Particular care should be taken to ensure that, when in use, the products do not endanger patients’ health or safety. |
(16) |
In support of transparency in Community legislation, certain information related to medical devices and their conformity with Directive 93/42/EEC, in particular information on registration, on vigilance reports and on certificates, should be available to any interested party and the general public. |
(17) |
To better coordinate the application and efficiency of national resources when applied to issues related to Directive 93/42/EEC, the Member States should cooperate with each other and at international level. |
(18) |
As design for patient safety initiatives play an increasing role in public health policy, it is necessary to expressly set out the need to consider ergonomic design in the essential requirements. In addition the level of training and knowledge of the user, such as in the case of a lay user, should be further emphasised within the essential requirements. The manufacturer should place particular emphasis on the consequences of misuse of the product and its adverse effects on the human body. |
(19) |
In the light of experience gained regarding activities of both the notified bodies and the authorities in the assessment of devices which require intervention of the appropriate authorities for medicines and human blood derivatives their duties and tasks should be clarified. |
(20) |
Taking account of the growing importance of software in the field of medical devices, be it as stand alone or as software incorporated in a device, validation of software in accordance with the state of the art should be an essential requirement. |
(21) |
In the light of the increased use of third Parties to carry out the design and manufacture of devices on behalf of the manufacturer, it is important that the manufacturer demonstrates that he applies adequate controls to the third party to continue to ensure the efficient operating of the quality system. |
(22) |
The classification rules are based on the vulnerability of the human body taking account of the potential risks associated with the technical design and manufacture of the devices. Explicit prior authorisation with regard to conformity, including an assessment of the design documentation, is required for Class III devices to be placed on the market. In performing its duties under the quality assurance and verification conformity assessment modules for all other classes of devices, it is essential and necessary for a notified body, in order to be assured of the compliance of the manufacturer with Directive 93/42/EEC, to review the design documentation for the medical device. The depth and extent of this review should be commensurate with the classification of the device, the novelty of the intended treatment, the degree of intervention, the novelty of the technology or construction materials, and the complexity of the design and/or technology. This review can be achieved by taking a representative example of design documentation of one or more type(s) of devices from those being manufactured. Further review(s), and in particular the assessment of changes to the design that could affect conformity with the essential requirements, should be part of the surveillance activities of the notified body. |
(23) |
It is necessary to remove the incoherence in the classification rules as a result of which invasive devices with respect to body orifices intended for connection to an active Class I medical device were not classified. |
(24) |
The measures necessary for the implementation of Directive 90/385/EEC and Directive 93/42/EEC should be adopted in accordance with Council Decision 1999/468/EC of 28 June 1999 laying down the procedures for the exercise of implementing powers conferred on the Commission (11). |
(25) |
In particular, power should be conferred on the Commission to adapt classification rules for medical devices, to adapt the means by which the information needed to use medical devices safely and properly may be set out, to determine conditions for making certain information publicly available, to adapt the provisions on clinical investigations set out in certain Annexes, to adopt particular requirements for placing certain medical devices on the market or putting them into service, and to take decisions to withdraw such devices from the market for reasons of protection of health or safety. Since those measures are of general scope and are designed to amend or supplement Directive 90/385/EEC and Directive 93/42/EEC by the modification or addition of non-essential elements, they must be adopted in accordance with the regulatory procedure with scrutiny provided for in Article 5a of Decision 1999/468/EC. |
(26) |
When, on imperative grounds of urgency, the normal time limits for the regulatory procedure with scrutiny cannot be complied with, the Commission should be able to use the urgency procedure provided for in Article 5a(6) of Decision 1999/468/EC for taking decisions on withdrawal of certain medical devices from the market and for the adoption of particular requirements for placing such devices on the market or putting them into service for reasons of protection of health or safety. |
(27) |
The Commission should give a mandate to CEN and/or Cenelec to specify technical requirements and a suitable specific label for phthalate-containing devices within 12 months after entry into force of this Directive. |
(28) |
Many Member States have established recommendations with the aim of reducing or limiting the use of medical devices containing critical phthalates on children, pregnant and nursing women and other patients at risk. To enable medical professionals to avoid such risks, devices which possibly release phthalates to the body of the patient should be labelled accordingly. |
(29) |
In accordance with the essential requirements on the design and manufacture of medical devices, manufacturers should avoid the use of substances that may possibly compromise the health of patients, in particular of substances which are carcinogenic, mutagenic or toxic to reproduction, and should, as appropriate, strive to develop alternative substances or products with a lower risk potential. |
(30) |
It should be clarified that alongside Directives 90/385/EEC and 93/42/EEC, in vitro diagnostic medical devices, which are the subject of Directive 98/79/EC of the European Parliament and of the Council of 27 October 1998 on in vitro diagnostic medical devices (12), should also be excluded from the scope of Directive 98/8/EC. |
(31) |
In accordance with point 34 of the Interinstitutional agreement on better law-making (13), Member States are encouraged to draw up, for themselves and in the interests of the Community, their own tables illustrating, as far as possible, the correlation between this Directive and the transposition measures, and to make them public. |
(32) |
Directives 90/385/EEC, 93/42/EEC and 98/8/EC should therefore be amended accordingly, |
HAVE ADOPTED THIS DIRECTIVE:
Article 1
Directive 90/385/EEC is hereby amended as follows:
1. |
Article 1 shall be amended as follows:
|
2. |
Article 2 shall be replaced by the following: ‘Article 2 Member States shall take all necessary steps to ensure that the devices may be placed on the market and/or put into service only if they comply with the requirements laid down in this Directive when duly supplied, properly implanted and/or properly installed, maintained and used in accordance with their intended purposes.’; |
3. |
Article 3 shall be replaced by the following: ‘Article 3 The active implantable medical devices referred to in Article 1(2)(c), (d) and (e), hereinafter referred to as “devices”, shall satisfy the essential requirements set out in Annex 1 which apply to them, account being taken of the intended purpose of the devices concerned. Where a relevant hazard exists, devices which are also machinery within the meaning of Article 2(a) of Directive 2006/42/EC of the European Parliament and of the Council of 17 May 2006 on machinery (*3) shall also meet the essential health and safety requirements set out in Annex I to that Directive to the extent to which those essential health and safety requirements are more specific than the essential requirements set out in Annex 1 to this Directive. |
4. |
in Article 4, paragraphs 1, 2 and 3 shall be replaced by the following: ‘1. Member States shall not create any obstacle to the placing on the market or the putting into service within their territory of devices complying with the provisions of this Directive and bearing the CE marking provided for in Article 12, which indicates that they have been the subject of an assessment of their conformity in accordance with Article 9. 2. Member States shall not create any obstacles to:
These devices shall not bear the CE marking. 3. At trade fairs, exhibitions, demonstrations, etc., Member States shall not create any obstacle to the showing of devices which do not conform to this Directive, provided that a visible sign clearly indicates that such devices do not conform and cannot be marketed or put into service until they have been made to comply by the manufacturer or his authorised representative.’; |
5. |
Article 5 shall be replaced by the following: ‘Article 5 1. Member States shall presume compliance with the essential requirements referred to in Article 3 in respect of devices which are in conformity with the relevant national standards adopted pursuant to the harmonised standards the references of which have been published in the Official Journal of the European Union; Member States shall publish the references of such national standards. 2. For the purposes of this Directive, reference to harmonised standards also includes the monographs of the European Pharmacopoeia notably on interaction between medicinal products and materials used in devices containing such medicinal products, the references of which have been published in the Official Journal of the European Union.’; |
6. |
Article 6 shall be amended as follows:
|
7. |
Article 8 shall be replaced by the following: ‘Article 8 1. Member States shall take the necessary steps to ensure that information brought to their knowledge regarding the incidents mentioned below involving a device is recorded and evaluated in a centralised manner:
2. Where a Member State requires medical practitioners or the medical institutions to inform the competent authorities of any incidents referred to in paragraph 1, it shall take the necessary steps to ensure that the manufacturer of the device concerned, or his authorised representative, is also informed of the incident. 3. After carrying out an assessment, if possible together with the manufacturer or his authorised representative, Member States shall, without prejudice to Article 7, immediately inform the Commission and the other Member States of measures that have been taken or are contemplated to minimise the recurrence of the incidents referred to in paragraph 1, including information on the underlying incidents. 4. The measures necessary for the implementation of this Article shall be adopted in accordance with the regulatory procedure referred to in Article 6(3).’; |
8. |
Article 9 shall be amended as follows:
|
9. |
Article 9a shall be replaced by the following: ‘Article 9a 1. A Member State shall submit a duly substantiated request to the Commission and ask it to take the necessary measures in the following situations:
Where measures are deemed necessary pursuant to the first subparagraph of this paragraph they shall be adopted in accordance with the regulatory procedure referred to in Article 6(3). 2. The Commission shall inform the Member States of the measures taken.’; |
10. |
Article 10 shall be amended as follows:
|
11. |
the following Articles shall be inserted: ‘Article 10a 1. Any manufacturer who, under his own name, places devices on the market in accordance with the procedure referred to in Article 9(2) shall inform the competent authorities of the Member State in which he has his registered place of business of the address of the registered place of business and the description of the devices concerned. Member States may request to be informed of all data allowing for the devices to be identified together with the label and the instructions for use when the devices are put into service within their territory. 2. Where a manufacturer who places a device on the market under his own name does not have a registered place of business in a Member State, he shall designate a single authorised representative in the European Union. For devices referred to in the first subparagraph of paragraph 1 the authorised representative shall inform the competent authority of the Member State in which he has his registered place of business of all details as referred to in paragraph 1. 3. The Member States shall on request inform the other Member States and the Commission of the details referred to in the first subparagraph of paragraph 1 given by the manufacturer or authorised representative. Article 10b 1. Regulatory data in accordance with this Directive shall be stored in a European databank accessible to the competent authorities to enable them to carry out their tasks relating to this Directive on a well-informed basis. The databank shall contain the following:
2. Data shall be forwarded in a standardised format. 3. The measures necessary for the implementation of paragraphs 1 and 2 of this Article, in particular paragraph 1(c), shall be adopted in accordance with the regulatory procedure referred to in Article 6(3). Article 10c Where a Member State considers in relation to a given product or group of products that, in order to ensure protection of health and safety and/or to ensure that public health requirements are observed, such products should be withdrawn from the market, or their placing on the market and putting into service should be prohibited, restricted or subjected to particular requirements, it may take any necessary and justified transitional measures. The Member State shall then inform the Commission and all the other Member States of the transitional measures, giving the reasons for its decision. The Commission shall, whenever possible, consult the interested Parties and the Member States. The Commission shall adopt its opinion, indicating whether the national measures are justified or not. The Commission shall inform all the Member States and the consulted interested Parties. When appropriate, the necessary measures designed to amend non-essential elements of this Directive, by supplementing it, relating to withdrawal from the market, prohibition of placing on the market and putting into service of a certain product or group of products or to restrictions or introduction of particular requirements therefor, shall be adopted in accordance with the regulatory procedure with scrutiny referred to in Article 6(4). On imperative grounds of urgency, the Commission may use the urgency procedure referred to in Article 6(5).’; |
12. |
Article 11 shall be amended as follows:
|
13. |
Article 13 shall be replaced by the following: ‘Article 13 Without prejudice to Article 7
Those provisions shall also apply where the CE marking has been affixed in accordance with the procedures in this Directive, but inappropriately, on products that are not covered by this Directive.’; |
14. |
Article 14 shall be amended as follows:
|
15. |
Article 15 shall be replaced by the following: ‘Article 15 1. Without prejudice to the existing national provisions and practices on medical confidentiality, Member States shall ensure that all the Parties involved in the application of this Directive are bound to observe confidentiality with regard to all information obtained in carrying out their tasks. This does not affect the obligations of Member States and notified bodies with regard to mutual information and the dissemination of warnings, nor the obligations of the persons concerned to provide information under criminal law. 2. The following information shall not be treated as confidential:
3. The measures designed to amend non-essential elements of this Directive, inter alia by supplementing it, relating to the determination of the conditions under which information other than that referred to in paragraph 2, and in particular concerning any obligation for manufacturers to prepare and make available a summary of the information and data related to the device, may be made publicly available shall be adopted in accordance with the regulatory procedure with scrutiny referred to in Article 6(4).’; |
16. |
the following Article shall be inserted: ‘Article 15a Member States shall take appropriate measures to ensure that the competent authorities of the Member States cooperate with each other and with the Commission and transmit to each other the information necessary to enable this Directive to be applied uniformly. The Commission shall provide for the organisation of an exchange of experience between the competent authorities responsible for market surveillance in order to coordinate the uniform application of this Directive. Without prejudice to the provisions of this Directive, cooperation may be part of initiatives developed at an international level.’; |
17. |
Annexes 1 to 7 shall be amended in accordance with Annex I to this Directive. |
Article 2
Directive 93/42/EEC is hereby amended as follows:
1. |
Article 1 shall be amended as follows:
|
2. |
in Article 3 the following paragraph shall be added: ‘Where a relevant hazard exists, devices which are also machinery within the meaning of Article 2(a) of Directive 2006/42/EC of the European Parliament and of the Council of 17 May 2006 on machinery (*10) shall also meet the essential health and safety requirements set out in Annex I to that Directive to the extent to which those essential health and safety requirements are more specific than the essential requirements set out in Annex I to this Directive. |
3. |
the second indent of Article 4(2) shall be replaced by the following:
|
4. |
in Article 6(1) the reference ‘83/189/EEC’ shall be replaced by the reference ‘98/34/EC (*11) |
5. |
Article 7 shall be replaced by the following: ‘Article 7 1. The Commission shall be assisted by the Committee set up by Article 6(2) of Directive 90/385/EEC, hereinafter referred to as “the Committee”. 2. Where reference is made to this paragraph, Articles 5 and 7 of Decision 1999/468/EC shall apply, having regard to the provisions of Article 8 thereof. The period laid down in Article 5(6) of Decision 1999/468/EC shall be set at three months. 3. Where reference is made to this paragraph, Article 5a(1) to (4) and Article 7 of Decision 1999/468/EC shall apply, having regard to the provisions of Article 8 thereof. 4. Where reference is made to this paragraph, Article 5a(1), (2), (4) and (6) and Article 7 of Decision 1999/468/EC shall apply, having regard to the provisions of Article 8 thereof.’; |
6. |
In Article 8 paragraph 2 shall be replaced by the following: ‘2. The Commission shall enter into consultation with the Parties concerned as soon as possible. Where, after such consultation, the Commission finds that:
|
7. |
In Article 9 paragraph 3 shall be replaced by the following: ‘3. Where a Member State considers that the classification rules set out in Annex IX require adaptation in the light of technical progress and any information which becomes available under the information system provided for in Article 10, it may submit a duly substantiated request to the Commission and ask it to take the necessary measures for adaptation of classification rules. The measures designed to amend non-essential elements of this Directive relating to adaptation of classification rules shall be adopted in accordance with the regulatory procedure with scrutiny referred to in Article 7(3).’; |
8. |
Article 10 shall be amended as follows:
|
9. |
Article 11 shall be amended as follows:
|
10. |
Article 12 shall be amended as follows:
|
11. |
The following Article shall be inserted: ‘Article 12a Reprocessing of medical devices The Commission shall, no later than 5 September 2010, submit a report to the European Parliament and to the Council on the issue of the reprocessing of medical devices in the Community. In the light of the findings of this report, the Commission shall submit to the European Parliament and to the Council any additional proposal it may deem appropriate in order to ensure a high level of health protection.’ |
12. |
Article 13 shall be replaced by the following: ‘Article 13 Decisions with regard to classification and derogation clause 1. A Member State shall submit a duly substantiated request to the Commission and ask it to take the necessary measures in the following situations:
The measures referred to in the first subparagraph of this paragraph shall, as appropriate, be adopted in accordance with the procedure referred to in Article 7(2). 2. The Commission shall inform the Member States of the measures taken.’; |
13. |
Article 14 shall be amended as follows:
|
14. |
Article 14a shall be amended as follows:
|
15. |
Article 14b shall be replaced by the following: ‘Article 14b Particular health monitoring measures Where a Member State considers, in relation to a given product or group of products, that, in order to ensure protection of health and safety and/or to ensure that public health requirements are observed, such products should be withdrawn from the market, or their placing on the market and putting into service should be prohibited, restricted or subjected to particular requirements, it may take any necessary and justified transitional measures. The Member State shall then inform the Commission and all other Member States, giving the reasons for its decision. The Commission shall, whenever possible, consult the interested Parties and the Member States. The Commission shall adopt its opinion, indicating whether the national measures are justified or not. The Commission shall inform all the Member States and the consulted interested Parties thereof. When appropriate, the necessary measures designed to amend non-essential elements of this Directive, relating to withdrawal from the market, prohibition of placing on the market and putting into service of a certain product or group of products or to restrictions or introduction of particular requirements in order for such products to be put on the market, shall be adopted in accordance with the regulatory procedure with scrutiny referred to in Article 7(3). On imperative grounds of urgency, the Commission may use the urgency procedure referred to in Article 7(4).’; |
16. |
Article 15 shall be amended as follows:
|
17. |
Article 16 shall be amended as follows:
|
18. |
in Article 18 point (a) shall be replaced by the following:
|
19. |
in Article 19(2), the words ‘established in the Community’ shall be deleted; |
20. |
Article 20 shall be replaced by the following: ‘Article 20 Confidentiality 1. Without prejudice to the existing national provisions and practices on medical confidentiality, Member States shall ensure that all the Parties involved in the application of this Directive are bound to observe confidentiality with regard to all information obtained in carrying out their tasks. This does not affect the obligation of Member States and notified bodies with regard to mutual information and the dissemination of warnings, nor the obligations of the persons concerned to provide information under criminal law. 2. The following information shall not be treated as confidential:
3. The measures designed to amend non-essential elements of this Directive, inter alia by supplementing it, relating to determination of the conditions under which other information may be made publicly available, and in particular for Class IIb and Class III devices to any obligation for manufacturers to prepare and make available a summary of the information and data related to the device, shall be adopted in accordance with the regulatory procedure with scrutiny referred to in Article 7(3).’; |
21. |
the following Article shall be inserted: ‘Article 20a Cooperation Member States shall take appropriate measures to ensure that the competent authorities of the Member States cooperate with each other and with the Commission and transmit to each other the information necessary to enable this Directive to be applied uniformly. The Commission shall provide for the organisation of an exchange of experience between the competent authorities responsible for market surveillance in order to coordinate the uniform application of this Directive. Without prejudice to the provisions of this Directive, cooperation may be part of initiatives developed at an international level.’; |
22. |
Annexes I to X shall be amended in accordance with Annex II to this Directive. |
Article 3
In Article 1(2) of Directive 98/8/EC the following point shall be added:
‘(s) |
Directive 98/79/EC of the European Parliament and of the Council of 27 October 1998 on in vitro diagnostic medical devices (*12). |
Article 4
1. Member States shall adopt and publish by 21 December 2008 the laws, regulations and administrative provisions necessary to comply with this Directive. They shall forthwith communicate to the Commission the text of those measures.
They shall apply those measures from 21 March 2010.
When Member States adopt those measures, they shall contain a reference to this Directive or be accompanied by such a reference on the occasion of their official publication. The methods of making such reference shall be laid down by Member States.
2. Member States shall communicate to the Commission the text of the main provisions of national law which they adopt in the field covered by this Directive.
Article 5
This Directive shall enter into force on the 20th day following its publication in the Official Journal of the European Union.
Article 6
This Directive is addressed to the Member States.
Done at Strasbourg, 5 September 2007.
For the European Parliament
The President
H.-G. PÖTTERING
For the Council
The President
M. LOBO ANTUNES
(1) OJ C 195, 18.8.2006, p. 14.
(2) Opinion of the European Parliament of 29 March 2007 (not yet published in the Official Journal) and Council Decision of 23 July 2007.
(3) OJ L 169, 12.7.1993, p. 1. Directive as last amended by Regulation (EC) No 1882/2003 of the European Parliament and of the Council (OJ L 284, 31.10.2003, p. 1).
(4) OJ L 311, 28.11.2001, p. 67. Directive as last amended by Regulation (EC) No 1901/2006 of the European Parliament and of the Council (OJ L 378, 27.12.2006, p. 1).
(6) OJ C 68 E, 18.3.2004, p. 85.
(7) OJ L 189, 20.7.1990, p. 17. Directive as last amended by Regulation (EC) No 1882/2003.
(8) OJ L 123, 24.4.1998, p. 1. Directive as last amended by Commission Directive 2007/20/EC (OJ L 94, 4.4.2007, p. 23).
(9) Directive 2000/70/EC of the European Parliament and of the Council of 16 November 2000 amending Council Directive 93/42/EEC as regards medical devices incorporating stable derivates of human blood or human plasma (OJ L 313, 13.12.2000, p. 22).
(10) OJ L 33, 8.2.2003, p. 30.
(11) OJ L 184, 17.7.1999, p. 23. Decision as amended by Decision 2006/512/EC (OJ L 200, 22.7.2006, p. 11).
(12) OJ L 331, 7.12.1998, p. 1. Directive as last amended by Regulation (EC) No 1882/2003.
ANNEX I
Annexes 1 to 7 to Directive 90/385/EEC shall be amended as follows:
1. |
Annex 1 shall be amended as follows:
|
2. |
Annex 2 shall be amended as follows:
|
3. |
Annex 3 shall be amended as follows:
|
4. |
Annex 4 shall be amended as follows:
|
5. |
Annex 5 shall be amended as follows:
|
6. |
Annex 6 shall be amended as follows:
|
7. |
Annex 7 shall be amended as follows:
|
(*1) OJ L 159, 29.6.1996, p. 1.
(*2) OJ L 180, 9.7.1997, p. 22.’;
(*3) Regulation (EC) No 726/2004 of the European Parliament and of the Council of 31 March 2004 laying down Community procedures for the authorisation and supervision of medicinal products for human and veterinary use and establishing a European Medicines Agency (OJ L 136, 30.4.2004, p. 1). Regulation as last amended by Regulation (EC) No 1901/2006.’;’
ANNEX II
Annexes I to X to Directive 93/42/EEC shall be amended as follows:
1. |
Annex I shall be amended as follows:
|
2. |
Annex II shall be amended as follows:
|
3. |
Annex III shall be amended as follows:
|
4. |
Annex IV shall be amended as follows:
|
5. |
Annex V shall be amended as follows:
|
6. |
Annex VI shall be amended as follows:
|
7. |
Annex VII shall be amended as follows:
|
8. |
Annex VIII shall be amended as follows:
|
9. |
Annex IX shall be amended as follows:
|
10. |
Annex X shall be amended as follows:
|
(*1) Regulation (EC) No 726/2004 of the European Parliament and of the Council of 31 March 2004 laying down Community procedures for the authorisation and supervision of medicinal products for human and veterinary use and establishing a European Medicines Agency (OJ L 136, 30.4.2004, p. 1). Regulation as last amended by Regulation (EC) No 1901/2006.’
(*2) OJ 196, 16.8.1967, p. 1. Directive as last amended by Directive 2006/121/EC of the European Parliament and of the Council (OJ L 396, 30.12.2006, p. 850).’
(*3) Commission Directive 2003/32/EC of 23 April 2003 introducing detailed specifications as regards the requirements laid down in Council Directive 93/42/EEC with respect to medical devices manufactured utilising tissues of animal origin (OJ L 105, 26.4.2003, p. 18).’ ’