EN
ANNEX I
GENERAL COMMON SPECIFICATIONS
Part I – Requirements for performance characteristics of devices covered by Annexes II to XIII
|
Performance characteristics
|
Requirement
|
|
All performance characteristics set out in Section 9.1, points (a) and (b), Section 9.3 and Section 9.4, point (a), of Annex I to Regulation (EU) 2017/746
|
1.The determination of performance characteristics shall be carried out in direct comparison with a state-of-the-art device. The device used for comparison shall be one bearing CE marking, if on the market at the time of the performance evaluation.
2.Devices used for determination of status of samples used in determination of performance characteristics shall be state-of-the-art devices bearing CE marking.
3.If discrepant results are identified as part of determination of performance characteristics, these results shall be resolved as far as possible, by one or more of the following:
–by evaluation of the discrepant sample in further devices,
–by use of an alternative method or marker,
–by a review of the clinical status and diagnosis of the patient,
–by the testing of follow-up samples.
4.The determination of performance characteristics shall be performed on a population equivalent to the European population.
|
|
Whole system failure rate
|
5.As part of the required risk analysis the whole system failure rate leading to false negative results shall be determined in repeat assays on low-positive specimens.
|
|
Analytical sensitivity and analytical specificity, interference
|
6.For devices intended for use with plasma the manufacturer shall verify the performance of the device using all anticoagulants which the manufacturer indicates for use with the device, for at least 50 plasma specimens (for devices intended for detection and/or quantification of infectious agents, 25 positive and 25 negative).
|
|
Analytical and diagnostic specificity, interference and cross-reactivity
|
7.The manufacturer shall select the potential interfering substances to be evaluated taking account of the composition of the reagents and configuration of the device.
|
|
Batch-to-batch consistency
|
8.For devices intended to detect antigens and antibodies, the manufacturer’s batch testing criteria shall ensure that every batch consistently identifies the relevant antigens, epitopes, and antibodies and is suitable for the claimed specimen types.
9.The manufacturer’s batch release testing for first-line assays, except those covered by Tables 1 and 2 of Annex XIV, shall include at least 100 specimens negative for the relevant analyte.
|
Part II – Requirements for performance characteristics of devices referred to in Annexes III to XIII
|
Performance characteristic
|
Requirement
|
|
Analytical and diagnostic sensitivity
|
10.Devices intended by the manufacturer for testing body fluids other than serum or plasma, e.g. urine, saliva, etc., shall meet the same requirements as serum or plasma devices. The manufacturer shall test samples from the same individuals in both the devices to be approved and in a respective serum or plasma device.
11.Devices for self-testing shall meet the same requirements as respective devices for professional use.
12.Positive specimens used in the performance evaluation shall be selected to reflect different stages of the respective disease(s), different antibody patterns, different genotypes, different subtypes, mutants, etc.
13.Seroconversion panels shall start with a negative bleed(s) and shall reflect narrow bleeding intervals as far as possible. Where this is not possible, manufacturers shall provide a justification in the performance evaluation report.
14.For devices intended by the manufacturer to be used with serum and plasma the performance evaluation must demonstrate serum to plasma equivalency. This shall be demonstrated for at least 25 positive donations.
15.For devices detecting or quantifying antigens or nucleic acids, the target antigen or gene respectively shall be specified in the instructions for use.
16.For devices detecting or quantifying antibodies against an infectious agent, the target antigen(s) of those antibodies shall be specified in the instructions for use.
|
|
Analytical and diagnostic specificity
|
17.Devices intended by the manufacturer for testing body fluids other than serum or plasma, e.g. urine, saliva, etc., shall meet the same requirements as serum or plasma devices. The performance evaluation shall test samples from the same individuals in both the devices to be approved and in a respective serum or plasma device.
18.Devices for self-testing shall meet the same requirements as respective devices for professional use.
19.Negative specimens used in a performance evaluation shall be defined so as to reflect the target population for which the device is intended, such as blood donors, hospitalised patients, pregnant women, etc.
20.Specificity shall be calculated using the frequency of repeatedly reactive (i.e. false positive) results in samples negative for the target marker.
21.For devices intended by the manufacturer to be used with serum and plasma the performance evaluation must demonstrate serum to plasma equivalency. This shall be demonstrated for at least 25 negative donations.
|
|
Analytical and diagnostic specificity, interference and cross-reactivity
|
22.The manufacturer shall include specimens such as, where applicable:
–specimens representing related infections,
–specimens from multipara, i.e. women who have had more than one pregnancy, or rheumatoid factor (RF) positive patients,
–specimens containing human antibodies to components of the expression system, for example anti-E. coli, or anti-yeast.
|
|
Performances obtained by lay persons
|
23.Relevant parts of the performance evaluation shall be carried out (or repeated) by appropriate lay persons to validate the operation of the device and the instructions for use. The lay persons selected for the performance evaluation shall be representative of the intended users groups.
|
ANNEX II
COMMON SPECIFICATIONS FOR DEVICES INTENDED FOR DETECTION OF BLOOD GROUP ANTIGENS IN THE ABO, RH, KELL, DUFFY AND KIDD BLOOD GROUP SYSTEMS
Scope
This Annex applies to devices intended for detection of blood group antigens in the ABO, Rh, Kell, Duffy and Kidd blood group systems.
Table 1 applies to performance evaluation of devices detecting blood group antigens in the ABO, Rh, Kell, Duffy and Kidd blood group systems.
Table 2 applies to manufacturer’s batch-to-batch consistency testing of reagents and reagent products to determine blood group antigens in the ABO, Rh, Kell, Duffy and Kidd blood group systems (test reagents, control materials).
Table 1. Performance evaluation of devices detecting blood group antigens in the ABO, Rh, Kell, Duffy and Kidd blood group systems
|
Reagent specificity
|
Number of tests per method claimed by the manufacturer
|
Total number of samples to be tested for a launch device
|
Total number of samples to be tested for a new formulation, or use of well-characterised reagents
|
General qualification criteria
|
Specific qualification criteria
|
Acceptance criteria
|
|
Anti-ABO1 (Anti-A), Anti-ABO2 (Anti-B), Anti-ABO3 (Anti-A,B)
|
≥500
|
≥3 000
|
≥1 000
|
Clinical samples: 10 % of the test population
Neonatal specimens: > 2 % of the test population
|
ABO samples shall include > 40 % A and B antigen positive samples which may include samples from group A, group B and group AB
|
All of the reagents shall show comparable performance to state-of-the-art CE marked devices with regard to claimed reactivity of the device.
For CE marked devices where the application or use has been changed or extended, further testing shall be carried out in accordance with the requirements outlined in column 2 above (“Number of tests per method claimed by the manufacturer”).
|
|
Anti-RH1 (Anti-D)
|
≥500
|
≥3 000
|
≥1 000
|
|
Performance evaluation of Anti-D reagents shall include tests against a range of weak RH1 (D) and partial RH1 (D) samples, depending on the intended use of the product.
Weak and/or partial D cells shall account for > 2 % of RH1 (D) positive samples.
|
|
|
Anti-RH2 (Anti-C), Anti-RH4 (Anti-c), Anti- RH3 (Anti-E)
|
≥100
|
≥1 000
|
≥200
|
|
|
|
|
Anti-RH5 (Anti-e)
|
≥100
|
≥500
|
≥200
|
|
|
|
|
Anti-KEL1 (Anti-K)
|
≥100
|
≥500
|
≥200
|
|
|
|
|
Anti-JK1 (Jka), Anti-JK2 (Jkb)
|
≥100
|
≥500
|
≥200
|
|
|
|
|
Anti-FY1 (Fya), Anti-FY2 (Fyb)
|
≥100
|
≥500
|
≥200
|
|
|
|
Note: Positive specimens used in the performance evaluation shall be selected to reflect variant and weak antigen expression.
Table 2. Manufacturer’s batch-to-batch consistency testing of reagents and reagent products to determine blood group antigens in the ABO, Rh, Kell, Duffy and Kidd blood group systems
1.Test reagents
|
Blood group reagents
|
Minimum number of control cells to be tested as part of specificity testing
|
Acceptance criteria
|
|
|
Positive reactions
|
|
Negative reactions
|
Each batch of reagent shall exhibit unequivocal positive or negative results by all techniques claimed by the manufacturer in accordance with the results obtained from the performance evaluation data.
|
|
|
A1
|
A2B
|
Ax
|
|
|
B
|
O
|
|
|
|
Anti-ABO1(Anti-A)
|
2
|
2
|
2
I
|
|
|
2
|
2
|
|
|
|
|
B
|
A1B
|
|
|
|
A1
|
O
|
|
|
|
Anti-ABO2(Anti-B)
|
2
|
2
|
|
|
|
2
|
2
|
|
|
|
|
A1
|
A2
|
Ax
|
B
|
|
O
|
|
|
|
|
Anti-ABO3(Anti-A,B)
|
2
|
2
|
21
|
2
|
|
4
|
|
|
|
|
|
R1r
|
R2r
|
WeakD
|
|
|
r’r
|
r”r
|
rr
|
|
|
Anti-RH1 (Anti-D)
|
2
|
2
|
21
|
|
|
1
|
1
|
1
|
|
|
|
R1R2
|
R1r
|
r’r
|
|
|
R2R2
|
r”r
|
rr
|
|
|
Anti-RH2 (Anti-C)
|
2
|
1
|
1
|
|
|
1
|
1
|
1
|
|
|
|
R1R2
|
R1r
|
r’r
|
|
|
R1R1
|
|
|
|
|
Anti-RH4 (Anti-c)
|
1
|
2
|
1
|
|
|
3
|
|
|
|
|
|
R1R2
|
R2r
|
r”r
|
|
|
R1R1
|
r’r
|
rr
|
|
|
Anti-RH3 (Anti-E)
|
2
|
1
|
1
|
|
|
1
|
1
|
1
|
|
|
|
R1R2
|
R2r
|
r”r
|
|
|
R2R2
|
|
|
|
|
Anti-RH5 (Anti-e)
|
2
|
1
|
1
|
|
|
3
|
|
|
|
|
|
Kk
|
|
|
|
|
kk
|
|
|
|
|
Anti-KEL1 (Anti-K)
|
4
|
|
|
|
|
3
|
|
|
|
|
|
Jk(a+b+)
|
|
|
|
|
Jk(a−b+)
|
|
|
|
|
Anti-JK1 (Anti-Jka)
|
4
|
|
|
|
|
3
|
|
|
|
|
|
Jk(a+b+)
|
|
|
|
|
Jk(a+b−)
|
|
|
|
|
Anti-JK2 (Anti-Jkb)
|
4
|
|
|
|
|
3
|
|
|
|
|
|
Fy(a+b+)
|
|
|
|
|
Fy(a−b+)
|
|
|
|
|
Anti-FY1 (Anti-Fya)
|
4
|
|
|
|
|
3
|
|
|
|
|
|
Fy(a+b+)
|
|
|
|
|
Fy(a+b−)
|
|
|
|
|
Anti-FY2 (Anti-Fyb)
|
4
|
|
|
|
|
3
|
|
|
|
Note: Polyclonal reagents shall be tested against a wider panel of cells to confirm specificity and exclude presence of unwanted contaminating antibodies.
2.Control materials (red cells)
The phenotype of red cells used in the control of blood typing reagents listed above shall be confirmed using established device.
ANNEX III
COMMON SPECIFICATIONS FOR DEVICES INTENDED FOR DETECTION OR QUANTIFICATION OF MARKERS OF HUMAN IMMUNODEFICIENCY VIRUS (HIV) INFECTION
Scope
1. This Annex applies to devices intended for detection or quantification of markers of human immunodeficiency virus (HIV) infection.
Table 1 applies to first-line assays for HIV-1/2 antibody (anti-HIV 1/2) and first-line combined antigen/antibody assays for HIV 1/2 (HIV 1/2 Ag/Ab) which are not rapid tests.
Table 2 applies to first-line assays for anti-HIV 1/2 and HIV 1/2 Ag/Ab which are rapid tests.
Table 3 applies to confirmatory assays for anti-HIV 1/2.
Table 4 applies to antigen tests for HIV-1 and HIV Ag/Ab assays.
Table 5 applies to for qualitative and quantitative NAT devices for HIV ribonucleic acid (RNA).
Table 6 applies to HIV-1/2 self-tests.
Definitions
2. For the purposes of this Annex, the following definitions apply:
(1)‘seroconversion HIV sample’ means:
–p24 antigen and/or HIV RNA positive, and
–recognised by the antibody first-line assays, and
–positive or indeterminate in confirmatory assays.
(2)‘early seroconversion HIV sample’ means:
–p24 antigen and/or HIV RNA positive, and
–not recognised by the antibody first-line assays, and
–indeterminate or negative in confirmatory assays.
Table 1. First-line assays: anti-HIV 1/2, HIV 1/2 Ag/Ab (requirements for antibody detection)
|
Performance characteristic
|
Specimen
|
Specimen numbers, features, use
|
Acceptance criteria
|
|
Diagnostic sensitivity
|
Positive specimens
|
≥400 HIV-1
≥100 HIV-2
including 40 non-B-subtypes
including 25 positive ‘same day’ fresh serum samples (≤ 1 day after sampling)
all available HIV/1 subtypes shall be represented by at least 3 samples per subtype
|
all true positive samples shall be identified as positive
|
|
|
Seroconversion panels
|
≥20 panels
≥10 further panels (at notified body or manufacturer)
at least 40 early seroconversion HIV samples shall be tested
|
diagnostic sensitivity during seroconversion shall represent the state of the art
all seroconversion HIV samples shall be identified as positive
|
|
Diagnostic specificity
|
Unselected blood donors (including first-time donors)
II
|
≥5 000
|
≥99,5%
|
|
|
Hospitalised patients
|
≥200
|
Potential limitations for specificity, if any, shall be identified
|
|
Cross-reactivity
|
Potentially cross-reacting specimens
|
≥100 in total
(such as RF+, from related virus infections, from pregnant women, subjects recently vaccinated against any infectious agent)
|
|
Table 2. Rapid tests: anti-HIV 1/2, HIV 1/2 Ag/Ab (requirements for antibody detection)
|
Performance characteristic
|
Specimen
|
Specimen numbers, features, use, acceptance criteria
|
Acceptance criteria
|
|
Diagnostic sensitivity
|
Positive specimens
|
≥400 HIV-1
≥100 HIV-2
including 40 non-B-subtypes
all available HIV/1 subtypes shall be represented by at least 3 samples per subtype
|
all true positive samples shall be identified as positive
|
|
|
Seroconversion panels
|
≥20 panels
≥10 further panels (at notified body or manufacturer)
at least 40 early seroconversion HIV samples shall be tested
|
diagnostic sensitivity during seroconversion shall represent the state of the art
all seroconversion HIV samples shall be identified as positive
|
|
Diagnostic specificity
|
Unselected blood donors (including first-time donors)
|
≥1 000
|
≥ 99 %
|
|
|
Hospitalised patients
|
≥200
|
Potential limitations for specificity, if any, shall be identified
|
|
Cross-reactivity
|
Potentially cross-reacting specimens
|
≥200 samples from pregnant women
≥100 other potentially cross-reacting specimens in total (e.g. RF+, from related infections)
|
|
Table 3. Confirmatory assays: anti-HIV 1/2
|
Performance characteristic
|
Specimen
|
Specimen numbers, features, use
|
Acceptance criteria
|
|
Diagnostic sensitivity
|
Positive specimens
|
≥200 HIV-1
≥100 HIV-2
Including different stages of infection and reflecting different antibody patterns
|
Identification as “confirmed positive” or “indeterminate”, not as “negative”
|
|
|
Seroconversion panels
|
≥15 seroconversion panels/low titre panels
≥40 early seroconversion HIV samples shall be tested
|
Diagnostic sensitivity during seroconversion shall represent the state of the art
All seroconversion HIV samples shall be identified as positive
|
|
Diagnostic specificity
|
Blood donors
|
≥200
|
No false positive results / no neutralisation
|
|
|
Hospitalised patients
|
≥200
|
|
|
Cross-reactivity
|
Potentially cross-reacting specimens
|
≥50 in total (including samples from pregnant women, samples with indeterminate results in other confirmatory assays)
|
|
Table 4. Antigen tests: HIV-1, HIV Ag/Ab (requirements for antigen detection)
|
Performance characteristic
|
Specimen
|
Specimen numbers, features, use
|
Acceptance criteria
|
|
Diagnostic sensitivity
|
Positive specimens
|
≥50 HIV-1 antigen positive
≥50 cell culture supernatants including different HIV-1 subtypes and HIV-2
|
all true positive samples shall be identified as positive (after neutralisation if applicable)
|
|
|
Seroconversion panels
|
≥20 seroconversion panels/low titre panels
≥40 early seroconversion HIV samples as referred to in Section 2(2) of this Annex shall be tested
|
diagnostic sensitivity during seroconversion shall represent the state of the art
all seroconversion HIV samples shall be identified as positive
|
|
Analytical sensitivity
|
First International Reference Reagent HIV-1 p24 Antigen, NIBSC code: 90/636
|
|
≤ 2 IU/ml
|
|
Diagnostic specificity
|
Blood donors
|
≥200
|
≥ 99,5 % after neutralisation or, if no neutralisation test available, after resolution of the sample status
|
|
|
Hospitalised patients
|
≥200
|
Potential limitations for specificity, if any, shall be identified
|
|
Cross-reactivity
|
Potentially cross-reacting specimens
|
≥50
|
|
Table 5. Qualitative and quantitative NAT devices for HIV RNA
1.For target sequence amplification devices, a functionality control for each specimen (internal control) shall reflect the state of the art. This control shall as far as possible be used throughout the whole process, i.e. extraction, amplification/hybridisation, detection.
2.Genotype and/or subtype detection shall be demonstrated by appropriate primer or probe design validation and shall also be validated by testing characterised genotyped samples.
3.Potential cross-reactivity of non-target nucleic acid sequences shall be analysed by appropriate primer or probe design validation and shall also be validated by testing selected samples
4.Results of quantitative NAT devices shall be traceable to international standards or calibrated reference materials, if available, and be expressed in international units utilised in the specific field of application.
5.Qualitative HIV NAT devices intended to be used to detect the presence of HIV in blood, blood components, cells, tissues or organs, or in any of their derivatives, in order to assess their suitability for transfusion, transplantation or cell administration shall be designed to detect both HIV-1 and HIV-2.
6.Qualitative HIV NAT devices, other than virus typing devices, shall be designed to compensate for the potential failure of a HIV-1 NAT target region by using two independent target regions.
|
Performance characteristic
|
Specimen
|
Specimen numbers, features, use
|
Acceptance criteria
|
|
Analytical sensitivity
|
WHO International Standard HIV-1 RNA; WHO International Standard HIV-2 RNA; or calibrated reference materials
|
NAT sensitivity and NAT LOD shall be validated by dilution series of reference materials, testing of replicates (minimal 24) at different analyte concentrations, including those with transition from positive to negative results in the respective NAT assay.
LOD shall be expressed as 95% positive cut-off value (IU/ml) after statistical analysis (e.g. Probit).
III
quantitative NAT: definition of lower, upper quantification limit, precision, accuracy,
‘linear’ measuring range, ‘dynamic range’.
Reproducibility at different concentration levels
|
According to the state of the art
|
|
HIV geno-/subtype sensitivity
|
all relevant genotypes/subtypes, preferably from international reference materials
potential substitutes for rare HIV subtypes (to be quantified by appropriate methods): cell culture supernatants; in vitro transcripts; plasmids
|
Qualitative NAT: at least 10 samples/genotype or subtype
Quantitative NAT: dilution series for demonstration of quantification efficiencies
|
According to the state of the art
|
|
Diagnostic sensitivity
|
Positive specimens reflecting the routine conditions of users (e.g. no pre-selection of specimens)
|
Quantitative NAT: ≥100
Comparative results with another NAT test system shall be generated in parallel
|
According to the state of the art
|
|
|
Seroconversion panels
|
Qualitative NAT: ≥10 panels
Comparative results with another NAT test system shall be generated in parallel
|
According to the state of the art
|
|
Diagnostic specificity
|
Blood donor specimens
|
Qualitative NAT: ≥500
Quantitative NAT: ≥100
|
According to the state of the art
|
|
Cross-reactivity
|
Potentially cross-reacting specimens
|
≥10 human retrovirus positive samples (e.g. HTLV)
|
|
|
Carry-over
|
High HIV RNA positive;
HIV RNA negative
|
At least five runs with alternating high-positive and negative specimens shall be performed during robustness studies. The virus titres of the high positive samples shall be representative of high virus titres occurring naturally.
|
According to the state of the art
|
|
Detection in relation to antibody status
|
HIV-RNA positives: anti-HIV negative, anti-HIV positive
|
Pre-seroconversion (anti-HIV negative) and post-seroconversion (anti-HIV positive) specimens
|
According to the state of the art
|
|
Whole system failure rate
|
HIV RNA low positive
|
≥100 HIV RNA low-positive specimens shall be tested. These specimens shall contain a virus concentration equivalent to three times the 95 % positive cut-off virus concentration.
|
≥99% positive
|
Table 6. Additional requirements for HIV-1/2 self-tests
|
Performance characteristic
|
Specimens
IV
|
Specimen number
|
|
Result interpretation
supervised
V
|
Interpretation of tests carried out with contrived specimens
VI
by lay users reflecting a range of results:
·non-reactive
·reactive
·weak reactive
VII
·invalid
|
≥ 100
|
|
Diagnostic sensitivity
|
lay users that are known positive
|
≥ 200
|
|
Diagnostic specificity
|
lay users that do not know their status
|
≥ 400
|
|
|
Lay users that are at high risk of acquiring the infection
|
≥ 200
|
ANNEX IV
COMMON SPECIFICATIONS FOR DEVICES INTENDED FOR DETECTION OR QUANTIFICATION OF MARKERS OF HUMAN T-CELL LYMPHOTROPIC VIRUS (HTLV) INFECTION
Scope
This Annex applies to devices intended for detection or quantification of markers of human T-cell lymphotropic virus (HTLV) infection.
Table 1 applies to first-line assays for antibodies against HTLV I or II (anti-HTLV I/II) which are not rapid tests.
Table 2 applies to first-line assays for anti HTLV I/II which are rapid tests.
Table 3 applies to confirmatory assays for anti-HTLV I/II.
Table 4 applies to NAT devices for HTLV I/II.
Table 1. First-line assays: anti-HTLV I/II
|
Performance characteristic
|
Specimen
|
Specimen numbers, features, use
|
Acceptance criteria
|
|
|
|
Diagnostic sensitivity
|
Positive specimens
|
≥300 HTLV-I
≥100 HTLV-II
including 25 positive ‘same day’ fresh serum samples (≤ 1 day after sampling)
|
all true positive samples shall be identified as positive
|
|
|
Seroconversion panels
|
To be defined when available
|
diagnostic sensitivity during seroconversion shall represent the state of the art, if applicable
|
|
Diagnostic specificity
|
Unselected blood donors (including first-time donors)
VIII
|
≥5 000
|
≥ 99,5%
|
|
|
Hospitalised patients
|
≥200
|
Potential limitations for specificity, if any, shall be identified
|
|
Cross-reactivity
|
Potentially cross-reacting specimens
|
≥100 in total
(e.g. RF+, from related virus infections, from pregnant women,)
|
|
Table 2. Rapid tests: anti HTLV I/II
|
Performance characteristic
|
Specimen
|
Specimen numbers, features, use, acceptance criteria
|
Acceptance criteria
|
|
Diagnostic sensitivity
|
Positive specimens
|
≥300 HTLV-I
≥100 HTLV-II
|
all true positive samples shall be identified as positive
|
|
|
Seroconversion panels
|
To be defined when available
|
diagnostic sensitivity during seroconversion shall represent the state of the art, if applicable
|
|
Diagnostic specificity
|
Unselected blood donors (including first-time donors)
|
≥1000
|
≥ 99%
|
|
|
Hospitalised patients
|
≥200
|
Potential limitations for specificity, if any, shall be identified
|
|
Cross-reactivity
|
Potentially cross-reacting specimens
|
≥200 samples from pregnant women
≥100 other potentially cross-reacting specimens in total (e.g. RF+, from related infections)
|
|
Table 3. Confirmatory assays: anti-HTLV I/II
|
Performance characteristic
|
Specimen
|
Specimen numbers, features, use, acceptance criteria
|
Acceptance criteria
|
|
Diagnostic sensitivity
|
Positive specimens
|
≥200 HTLV I
≥100 HTLV II
|
Identification as “confirmed positive” or “indeterminate”, not as “negative”
|
|
|
Seroconversion panels
|
To be defined when available
|
diagnostic sensitivity during seroconversion shall represent the state of the art, if applicable
|
|
Diagnostic specificity
|
Blood donors
|
≥200
|
No false positive results
|
|
|
Hospitalised patients
|
≥200
|
|
|
Cross-reactivity
|
Potentially cross-reacting specimens
|
≥50 in total (including samples from pregnant women, samples with indeterminate results in other confirmatory assays)
|
|
Table 4. NAT devices for HTLV I/II
1.For target sequence amplification devices, a functionality control for each specimen (internal control) shall reflect the state of the art. This control shall as far as possible be used throughout the whole process, i.e. extraction, amplification/hybridisation, detection.
2.Genotype and/or subtype detection shall be demonstrated by appropriate primer or probe design validation and shall also be validated by testing characterised genotyped samples.
3.Potential cross-reactivity of non-target nucleic acid sequences shall be analysed by appropriate primer or probe design validation and shall also be validated by testing selected samples.
4.Results of quantitative NAT devices shall be traceable to international standards or calibrated reference materials, if available, and be expressed in international units utilised in the specific field of application.
|
Performance characteristic
|
Specimen
|
Specimen numbers, features, use, acceptance criteria
|
Acceptance criteria
|
|
Analytical sensitivity
|
International reference preparations
|
NAT sensitivity and NAT LOD shall be validated by dilution series of reference materials, testing of replicates (minimal 24) at different analyte concentrations, including those with transition from positive to negative results in the respective NAT assay.
LOD shall be expressed as 95% positive cut-off value (IU/ml) after statistical analysis (e.g. Probit).
IX
quantitative NAT: definition of lower, upper quantification limit, precision, accuracy,
‘linear’ measuring range, ‘dynamic range’.
Reproducibility at different concentration levels
|
According to the state of the art
|
|
HTLV I and HTLV II genotype sensitivity
|
all relevant genotypes, preferably from international reference materials
potential substitutes for rare HTLV genotypes (to be quantified by appropriate methods): cell culture supernatants; in vitro transcripts; plasmids
|
Qualitative NAT: at least 10 samples/genotype or subtype
Quantitative NAT: dilution series for demonstration of quantification efficiencies
|
According to the state of the art
|
|
Diagnostic specificity
|
Blood donor specimens
|
Qualitative NAT: ≥500
Quantitative NAT: ≥100
|
According to the state of the art
|
|
Cross-reactivity
|
Potentially cross-reacting specimens
|
≥10 human retrovirus positive samples (e.g. HIV-1, HIV-2)
|
According to the state of the art
|
|
Carry-over
|
High HTLV RNA positive;
HTLV RNA negative
|
At least five runs with alternating high-positive and negative specimens shall be performed during robustness studies. The virus titres of the high positive samples shall be representative of high virus titres occurring naturally.
|
According to the state of the art
|
|
Detection in relation to antibody status
|
HTLV-RNA positives: anti-HTLV negative, anti-HTLV positive
|
Pre-seroconversion (anti-HTLV negative) and post-seroconversion (anti-HTLV positive) specimens
|
According to the state of the art
|
|
Whole system failure rate
|
HTLV RNA low positive
|
≥100 HTLV RNA low-positive specimens shall be tested. These specimens shall contain a virus concentration equivalent to three times the 95 % positive cut-off virus concentration.
|
≥99% positive
|
ANNEX V
COMMON SPECIFICATIONS FOR DEVICES INTENDED FOR DETECTION OR QUANTIFICATION OF MARKERS OF HEPATITIS C VIRUS (HCV) INFECTION
Scope
This Annex applies to devices intended for detection or quantification of markers of hepatitis C virus (HCV) infection.
Table 1 applies to first-line assays for anti-HCV antibodies (anti-HCV) and combined antigen/antibody tests for HCV (HCV Ag/Ab) which are not rapid tests.
Table 2 applies to first-line assays for anti-HCV and HCV Ag/Ab which are rapid tests.
Table 3 applies to confirmatory and supplemental assays for anti-HCV.
Table 4 applies to HCV antigen tests and HCV Ag/Ab.
Table 5 applies to qualitative and quantitative NAT devices for HCV RNA.
Table 6 applies to HCV self-tests.
Table 7 applies to HCV serotyping and genotyping assays.
Table 1. First-line assays: anti-HCV, HCV Ag/Ab (requirements for antibody detection)
|
Performance characteristic
|
Specimen
|
Specimen numbers, features, use
|
Acceptance criteria
|
|
Diagnostic sensitivity
|
Positive specimens
|
≥400
Including samples from different stages of infection and reflecting different antibody patterns
HCV genotype 1-4: > 20 samples per genotype (including non-a subtypes of genotype 4); HCV genotypes 5 and 6: > 5 samples each;
including 25 positive ‘same day’ fresh serum samples (≤ 1 day after sampling)
|
all true positive samples shall be identified as positive
|
|
|
Seroconversion panels
|
≥20 panels
≥10 further panels (at notified body or manufacturer)
HCV seroconversion panels for the evaluation of HCV antigen and antibody combined tests (HCV Ag/Ab) shall start with one or more negative bleeds and comprise panel members from early HCV infection (HCV core antigen and/or HCV RNA positive but anti-HCV negative).
|
diagnostic sensitivity during seroconversion shall represent the state of the art
HCV Ag/Ab tests shall demonstrate enhanced sensitivity in early HCV infection when compared to HCV antibody only tests.
|
|
Diagnostic specificity
|
Unselected blood donors (including first-time donors)
X
|
≥5 000
|
≥99,5%
|
|
|
Hospitalised patients
|
≥200
|
Potential limitations for specificity, if any, shall be identified
|
|
Cross-reactivity
|
Potentially cross-reacting specimens
|
≥100 in total
(e.g. RF+, from related virus infections, from pregnant women,)
|
|
Table 2. Rapid tests: anti-HCV, HCV Ag/Ab (requirements for antibody detection)
|
Performance characteristic
|
Specimen
|
Specimen numbers, features, use, acceptance criteria
|
Acceptance criteria
|
|
Diagnostic sensitivity
|
Positive specimens
|
≥400
including samples from different stages of infection and reflecting different antibody patterns.
HCV genotype 1-4: > 20 samples per genotype (including non-a subtypes of genotype 4); HCV genotypes 5 and 6: > 5 samples each;
|
all true positive samples shall be identified as positive
|
|
|
Seroconversion panels
|
≥20 panels
≥10 further panels (at notified body or manufacturer)
HCV seroconversion panels for the evaluation of HCV antigen and antibody combined tests (HCV Ag/Ab) shall start with one or more negative bleeds and comprise panel members from early HCV infection (HCV core antigen and/or HCV RNA positive but anti-HCV negative).
|
diagnostic sensitivity during seroconversion shall represent the state of the art
HCV Ag/Ab tests shall demonstrate enhanced sensitivity in early HCV infection when compared to HCV antibody only tests.
|
|
Diagnostic specificity
|
Unselected blood donors (including first-time donors)1
|
≥1 000
|
≥ 99 %
|
|
|
Hospitalised patients
|
≥200
|
Potential limitations for specificity, if any, shall be identified
|
|
Cross-reactivity
|
Potentially cross-reacting specimens
|
≥200 samples from pregnant women
≥100 other cross-reacting specimens in total (e.g. RF+, from related infections)
|
|
Table 3. Confirmatory and supplemental assays: anti-HCV
|
Performance characteristic
|
Specimen
|
Specimen numbers, features, use, acceptance criteria
|
Acceptance criteria
|
|
Diagnostic sensitivity
|
Positive specimens
|
≥300
Including specimens from different stages of infection and reflecting different antibody patterns.
HCV genotypes 1 – 4: > 20 specimens (including non-a subtypes of genotype 4; HCV genotype 5: > 5 specimens; HCV genotype 6: as far as available
|
identification as “confirmed positive” or “indeterminate”, not as “negative”
|
|
|
Seroconversion panels
|
≥15 seroconversion panels/low titre panels
|
diagnostic sensitivity during seroconversion shall represent the state of the art
|
|
Diagnostic specificity
|
Blood donors
|
≥200
|
No false positive results/ no neutralisation
|
|
|
Hospitalised patients
|
≥200
|
|
|
Cross-reactivity
|
Potentially cross-reacting specimens
|
≥50 in total (including samples from pregnant women, samples with indeterminate results in other confirmatory assays)
|
|
Table 4. Antigen tests: HCV antigen, HCV Ag/Ab (requirements for antigen detection)
|
Performance characteristic
|
Specimen
|
Specimen numbers, features, use, acceptance criteria
|
Acceptance criteria
|
|
Diagnostic sensitivity
|
Positive specimens
|
≥25 HCV core antigen and/or HCV RNA positive but anti-HCV negative specimens, comprising HCV genotypes 1-6 (if a genotype is not available, a justification shall be made)
|
all true positive specimens shall be identified as positive
|
|
|
Seroconversion panels
|
≥20 seroconversion panels/low titre panels
HCV seroconversion panels for the evaluation of HCV antigen and antibody combined tests shall start with one or more negative bleeds and comprise panel members from early HCV infection (HCV core antigen and/or HCV RNA positive but anti-HCV negative).
|
diagnostic sensitivity during seroconversion shall represent the state of the art
HCV antigen and antibody combined tests shall demonstrate enhanced sensitivity in early HCV infection when compared to HCV antibody only tests.
|
|
Analytical sensitivity
|
WHO International Standard HCV core (PEI 129096/12)
|
Dilution series
|
|
|
Diagnostic specificity
|
Blood donors
|
≥200
|
≥ 99,5 % after neutralisation or, if no neutralisation test available, after resolution of the sample status
|
|
|
Hospitalised patients
|
≥200
|
Potential limitations for specificity, if any, shall be identified
|
|
Cross-reactivity
|
Potentially cross-reacting specimens
|
≥50
|
|
Table 5. Qualitative and quantitative NAT devices for HCV RNA
1.For target sequence amplification devices, a functionality control for each specimen (internal control) shall reflect the state of the art. This control shall as far as possible be used throughout the whole process, i.e. extraction, amplification/hybridisation, detection.
2.Genotype and/or subtype detection shall be demonstrated by appropriate primer or probe design validation and shall also be validated by testing characterised genotyped samples.
3.Potential cross-reactivity of non-target nucleic acid sequences shall be analysed by appropriate primer or probe design validation and shall also be validated by testing selected samples.
4.Results of quantitative NAT devices shall be traceable to international standards or calibrated reference materials, if available, and be expressed in international units utilised in the specific field of application.
|
Performance characteristic
|
Specimen
|
Specimen numbers, features, use, acceptance criteria
|
Acceptance criteria
|
|
Analytical sensitivity
|
WHO International Standard HCV RNA(or calibrated reference materials)
|
NAT sensitivity and NAT LOD shall be validated by dilution series of reference materials, testing of replicates (minimal 24) at different analyte concentrations, including those with transition from positive to negative results in the respective NAT assay.
LOD shall be expressed as 95% positive cut-off value (IU/ml) after statistical analysis (e.g. Probit).
XI
quantitative NAT: definition of lower, upper quantification limit, precision, accuracy,
‘linear’ measuring range, ‘dynamic range’.
Reproducibility at different concentration levels
|
According to the state of the art
|
|
HCV genotype sensitivity
|
all relevant genotypes/subtypes, preferably from international reference materials
potential substitutes for rare HCV genotypes (to be quantified by appropriate methods): in vitro transcripts; plasmids
|
Qualitative NAT: ≥10 samples/genotype or subtype
Quantitative NAT: dilution series for demonstration of quantification efficiencies
|
According to the state of the art
|
|
Diagnostic sensitivity
|
Positive specimens reflecting the routine conditions of users (e.g. no pre-selection of specimens)
|
Quantitative NAT: ≥100
Comparative results with another NAT test system shall be generated in parallel
|
According to the state of the art
|
|
|
Seroconversion panels
|
Qualitative NAT: ≥10 panels
Comparative results with another NAT test system shall be generated in parallel
|
According to the state of the art
|
|
Diagnostic specificity
|
Blood donor specimens
|
Qualitative NAT: ≥500
Quantitative NAT: ≥100
|
According to the state of the art
|
|
Cross-reactivity
|
Potentially cross-reacting specimens
|
>10 human flavivirus (e.g. HGV, YFV) positive samples
|
According to the state of the art
|
|
Carry-over
|
High HCV RNA positive;
HCV RNA negative
|
At least five runs with alternating high-positive and negative specimens shall be performed during robustness studies. The virus titres of the high positive samples shall be representative of high virus titres occurring naturally.
|
According to the state of the art
|
|
Detection in relation to antibody status
|
HCV RNA positives: anti-HCV negative, anti-HCV positive
|
Pre-seroconversion (anti-HCV negative) and post-seroconversion (anti-HCV positive) specimens
|
According to the state of the art
|
|
Whole system failure rate
|
HCV RNA low positive
|
≥100 HCV RNA low-positive specimens shall be tested. These specimens shall contain a virus concentration equivalent to three times the 95 % positive cut-off virus concentration.
|
≥99% positive
|
Table 6. Additional requirements for HCV self-tests
|
Performance characteristic
|
Specimens
XII
|
Number of specimens
|
|
Result interpretation
supervised
XIII
|
Interpretation of tests carried out with contrived specimens
XIV
by lay users reflecting a range of results:
·non-reactive
·reactive
·weak reactive
XV
·invalid
|
≥ 100
|
|
Diagnostic sensitivity
|
lay users that are known positive
|
≥ 200
|
|
Diagnostic specificity
|
lay users that do not know their status
|
≥ 400
|
|
|
lay users that are at high risk of acquiring the infection
|
≥ 200
|
Table 7. HCV serotyping and genotyping assays
1.For target sequence amplification devices, a functionality control for each specimen (internal control) shall reflect the state of the art. This control shall as far as possible be used throughout the whole process, i.e. extraction, amplification/hybridisation, detection.
2.Genotype and/or subtype detection shall be demonstrated by appropriate primer or probe design validation and shall also be validated by testing characterised genotyped samples.
3.Potential cross-reactivity of non-target nucleic acid sequences shall be analysed by appropriate primer or probe design validation and shall also be validated by testing selected samples.
4.Results of quantitative NAT devices shall be traceable to international standards or calibrated reference materials, if available, and be expressed in international units utilised in the specific field of application.
|
Performance characteristic
|
Specimen
|
Specimen numbers, features, use, acceptance criteria
|
Acceptance criteria
|
|
Diagnostic sensitivity
|
Positive specimens
|
≥200
Including specimens from different stages of infection and reflecting different antibody patterns
HCV genotype 1-6: > 20 specimens per genotype
|
≥ 95 % agreement between serotyping and genotyping;
> 95 % agreement between genotyping and sequencing
|
|
Diagnostic specificity
|
Negative specimens
|
≥100
|
|
ANNEX VI
COMMON SPECIFICATIONS FOR DEVICES INTENDED FOR DETECTION OR QUANTIFICATION OF MARKERS OF HEPATITIS B VIRUS (HBV)INFECTION
Scope
This Annex applies to devices intended for detection or quantification of markers of hepatitis B virus (HBV) infection.
Table 1 applies to first-line assays for hepatitis B surface antigen (HBsAg), and for antibodies against hepatitis B core antigen (anti-HBc) which are not rapid tests.
Table 2 applies to first-line assays for HBsAg and anti-HBc which are rapid tests.
Table 3 applies to confirmatory assays for HBsAg.
Table 4 applies to assays for the hepatitis B virus markers: hepatitis B surface antibodies (anti-HBs), IgM antibody against the hepatitis B core antigen (anti-HBc IgM), antibodies against the hepatitis Be antigen (anti-HBe) and hepatitis Be antigen (HBeAg).
Table 5 applies to qualitative and quantitative NAT for HBV deoxyribonucleic acid (DNA).
Table 6 applies to HBV self-tests.
Table 1. First-line assays: HBsAg, anti-HBc
|
Performance characteristic
|
Specimen
|
Specimen numbers, features, use
|
Acceptance criteria
|
|
Diagnostic sensitivity
|
Positive specimens
|
≥400
anti-HBc: including evaluation of different HBV markers
HBsAg: including different HBV genotypes / subtypes / mutants
anti-HBc or HBsAg: including 25 positive ‘same day’ fresh serum (≤ 1 day after sampling)
|
Overall performance shall be at least equivalent to that of the state of the art device referred to in Annex I (3).
|
|
|
Seroconversion panels
|
HBsAg assays:
≥20 panels
≥10 further panels (at notified body or manufacturer)
antiHBc assays:
to be defined when available
|
diagnostic sensitivity during seroconversion shall represent the state of the art (for antiHBc this shall be the case if applicable)
|
|
Analytical sensitivity
|
WHO Third International Standard HBsAg (subtypes ayw1/adw2, HBV genotype B4, NIBSC code: 12/226)
|
|
For HBsAg assays: <0,130 IU/ml
|
|
Diagnostic specificity
|
Unselected blood donors (including first-time donors)
XVI
|
≥5 000
|
≥99,5%
|
|
|
Hospitalised patients
|
≥200
|
Potential limitations for specificity, if any, shall be identified
|
|
Cross-reactivity
|
Potentially cross-reacting specimens
|
≥100 in total
(e.g. RF+, from related virus infections, from pregnant women,)
|
|
Table 2. Rapid tests: HBsAg, anti-HBc
|
Performance characteristic
|
Specimen
|
Specimen numbers, features, use
|
Acceptance criteria
|
|
Diagnostic sensitivity
|
Positive specimens
|
≥400
including evaluation of different HBV markers
including different HBV genotypes / subtypes / mutants
|
Overall performance shall be at least equivalent to that of state of the art device referred to in Annex I (3)
|
|
|
Seroconversion panels
|
HBsAg assays:
≥20 panels
≥10 further panels (at notified body or manufacturer)
Anti-HBc assays:
to be defined when available
|
diagnostic sensitivity during seroconversion shall represent the state of the art (for anti-HBc this shall be the case if applicable)
|
|
Diagnostic specificity
|
Unselected blood donors (including first-time donors)
|
≥1 000
|
HBsAg assays: ≥ 99 %
anti-HBc assays: ≥ 99 %
|
|
|
Hospitalised patients
|
≥200
|
Potential limitations for specificity, if any, shall be identified
|
|
Cross-reactivity
|
Potentially cross-reacting specimens
|
≥200 samples from pregnant women
≥100 other potentially cross-reacting specimens in total (e.g. RF+, from related infections)
|
|
Table 3. Confirmatory assays: HBsAg
|
Performance characteristic
|
Specimen
|
Specimen numbers, features, use, acceptance criteria
|
Acceptance criteria
|
|
Diagnostic sensitivity
|
Positive specimens
|
≥300
Including samples from different stages of infection
Including 20 ‘high positive’ samples (>26 IU/ml); 20 samples in the cut-off range
|
Correct identification as positive (or indeterminate), not negative
|
|
|
Seroconversion panels
|
≥15 seroconversion panels/low titre panels
|
diagnostic sensitivity during seroconversion shall represent the state of the art
|
|
Analytical sensitivity
|
Third International Standard for HBsAg, subtypes ayw1/adw2, HBV genotype B4, NIBSC code: 12/226
|
|
|
|
Diagnostic specificity
|
Negative specimens
|
≥10 false positives as available from the performance evaluation of the first-line assay
|
No false positive results/ no neutralisation
|
|
Cross-reactivity
|
Potentially cross-reacting specimens
|
≥50
|
|
Table 4. Assays for the HBV markers: anti-HBs, anti-HBc IgM, anti-HBe, HBeAg
|
Performance characteristic
|
|
Anti-HBs
|
Anti-HBc IgM
|
Anti-HBe
|
HBeAg
|
Acceptance criteria
|
|
Diagnostic sensitivity
|
Positive specimens
|
≥100 vaccinees
≥100 naturally infected persons
|
≥200
Including samples from different stages of infection (acute/chronic, etc.)
|
≥200
Including samples from different stages of infection (acute/chronic, etc.)
|
≥200
Including samples from different stages of infection (acute/chronic, etc.)
|
≥ 98 %
(for Anti-HBc IgM: applicable only on samples from acute infection stage)
|
|
|
Seroconversion panels
|
10 follow-ups or anti-HBs seroconversions
|
When available
|
When available
|
When available
|
Diagnostic sensitivity during seroconversion shall represent the state of the art (for Anti-HBc IgM, Anti-HBe, HBeAg this shall be the case if applicable)
|
|
Analytical sensitivity
|
Standards
|
WHO 2nd International Standard for anti-hepatitis B surface antigen (anti-HBs) immunoglobulin, human NIBSC code: 07/164
|
|
WHO 1st International Standard Anti‑hepatitis B virus e antigen (anti‑HBe), PEI code 129095/12
|
WHO 1st International Standard for Hepatitis B Virus e Antigen (HBeAg) PEI code 129097/12 HBe
|
Anti-HBs: < 10 mIU/ml
|
|
Diagnostic specificity
|
Negative specimens
|
≥500
Including clinical samples
≥50 potentially interfering samples
|
≥200 blood donations
≥200 clinical samples
≥50 potentially interfering samples
|
≥200 blood donations
≥200 clinical samples
≥50 potentially interfering samples
|
≥200 blood donations
≥200 clinical samples
≥50 potentially interfering samples
|
≥ 98 %
|
Table 5. Qualitative and quantitative NAT devices for HBV DNA
1.For target sequence amplification devices, a functionality control for each specimen (internal control) shall reflect the state of the art. This control shall as far as possible be used throughout the whole process, i.e. extraction, amplification/hybridisation, detection.
2.Genotype and/or subtype detection shall be demonstrated by appropriate primer or probe design validation and shall also be validated by testing characterised genotyped samples.
3.Potential cross-reactivity of non-target nucleic acid sequences shall be analysed by appropriate primer or probe design validation and shall also be validated by testing selected samples.
4.Results of quantitative NAT devices shall be traceable to international standards or calibrated reference materials, if available, and be expressed in international units utilised in the specific field of application.
|
Performance characteristic
|
Specimen
|
Specimen numbers, features, use
|
Acceptance criteria
|
|
Analytical sensitivity
|
WHO International Standard HBV DNA (or calibrated reference materials)
|
NAT sensitivity and NAT LOD shall be validated by dilution series of reference materials, testing of replicates (minimal 24) at different analyte concentrations, including those with transition from positive to negative results in the respective NAT assay.
LOD shall be expressed as 95% positive cut-off value (IU/ml) after statistical analysis (e.g. Probit).
XVII
quantitative NAT: definition of lower, upper quantification limit, precision, accuracy,
‘linear’ measuring range, ‘dynamic range’. Reproducibility at different concentration levels
|
According to the state of the art
|
|
HBV genotype sensitivity
|
WHO International Reference Panel HBV DNA (HBV genotypes)
all relevant genotypes/subtypes, preferably from international reference materials
potential substitutes for rare HBV genotypes (to be quantified by appropriate methods): plasmids; synthetic DNA
|
Qualitative NAT: at least 10 samples/genotype or subtype
Quantitative NAT: dilution series for demonstration of quantification efficiencies
|
According to the state of the art
|
|
Diagnostic sensitivity
|
Positive specimens reflecting the routine conditions of users (e.g. no pre-selection of specimens)
|
Quantitative NAT: ≥100
Comparative results with another NAT test system shall be generated in parallel
|
According to the state of the art
|
|
|
Seroconversion panels
|
Qualitative NAT: ≥10 panels
Comparative results with another NAT test system shall be generated in parallel
|
According to the state of the art
|
|
Diagnostic specificity
|
Blood donor specimens
|
Qualitative NAT: ≥500
Quantitative NAT: ≥100
|
According to the state of the art
|
|
Cross-reactivity
|
Potentially cross-reacting specimens
|
|
According to the state of the art
|
|
Carry-over
|
High HBV DNA positive;
HBV DNA negative
|
At least five runs with alternating high-positive and negative specimens shall be performed during robustness studies. The high positive samples shall comprise samples with naturally occurring high virus titres.
|
According to the state of the art
|
|
Detection in relation to antibody status
|
HBV DNA positives: anti-HBV negative, anti-HBV positive
|
Pre-seroconversion (anti-HBV negative) and post-seroconversion (anti-HBV positive) specimens
|
According to the state of the art
|
|
Whole system failure rate
|
HBV DNA low positive
|
≥100 HBV DNA low-positive specimens shall be tested. These specimens shall contain a virus concentration equivalent to three times the 95 % positive cut-off virus concentration.
|
≥99% positive
|
Table 6. Additional requirements for HBV self-tests
|
Performance characteristic
|
Specimens
XVIII
|
Numbers of specimens
|
|
Result interpretation
supervised
XIX
|
Interpretation of tests carried out with contrived specimens
XX
by lay users reflecting a range of results:
·non-reactive
·reactive
·weak reactive
XXI
·invalid
|
≥100
|
|
Diagnostic sensitivity
|
lay users that are known positive
|
≥200
|
|
Diagnostic specificity
|
lay users that do not know their status
|
≥400
|
|
|
lay users that are at high risk of acquiring the infection
|
≥200
|
ANNEX VIII
COMMON SPECIFICATIONS FOR DEVICES INTENDED FOR DETECTION OR QUANTIFICATION OF MARKERS OF HEPATITIS D VIRUS (HDV) INFECTION
Scope
This Annex applies to devices intended for detection or quantification of markers of hepatitis D virus (HDV) infection.
Table 1 applies to devices intended for detection (including confirmation) or quantification of the following hepatitis D virus markers: antibodies against hepatitis D virus (anti-HDV), IgM antibodies against hepatitis D virus (anti-HDV IgM), the delta antigen.
Table 2 applies to qualitative and quantitative NAT for HDV RNA.
Table 1. Assays for HDV markers: anti-HDV, anti-HDV IgM, delta antigen
|
Performance characteristic
|
|
Anti-HDV
|
Anti-HDV IgM
|
Delta antigen
|
Acceptance criteria
|
|
Diagnostic sensitivity
|
Positive specimens
|
≥100
Specifying markers of HBV coinfection
|
≥50
Specifying markers of HBV coinfection
|
≥10
Specifying markers of HBV coinfection
|
≥ 98 %
|
|
Diagnostic specificity
|
Negative specimens
|
≥200
Including clinical samples
≥50 potentially interfering samples
|
≥200
Including clinical samples
≥50 potentially interfering samples
|
≥200
Including clinical samples
≥50 potentially interfering samples
|
≥ 98 %
|
Table 2. Qualitative and quantitative NAT devices for HDV RNA
1.For target sequence amplification devices, a functionality control for each specimen (internal control) shall reflect the state of the art. This control shall as far as possible be used throughout the whole process, i.e. extraction, amplification/hybridisation, detection.
2.Genotype and/or subtype detection shall be demonstrated by appropriate primer or probe design validation and shall also be validated by testing characterised genotyped samples.
3.Potential cross-reactivity of non-target nucleic acid sequences shall be analysed by appropriate primer or probe design validation and shall also be validated by testing selected samples.
4.Results of quantitative NAT devices shall be traceable to international standards or calibrated reference materials, if available, and be expressed in international units utilised in the specific field of application.
|
Performance characteristic
|
Specimen
|
Specimen numbers, features, use
|
Acceptance criteria
|
|
Analytical sensitivity
|
1st WHO International Standard HDV RNA, PEI code 7657/12
|
NAT sensitivity and NAT LOD shall be validated by dilution series of reference materials, testing of replicates (minimal 24) at different analyte concentrations, including those with transition from positive to negative results in the respective NAT assay.
LOD shall be expressed as 95% positive cut-off value (IU/ml) after statistical analysis (e.g. Probit).
XXII
quantitative NAT: definition of lower, upper quantification limit, precision, accuracy,
‘linear’ measuring range, ‘dynamic range’. Reproducibility at different concentration levels
|
According to the state of the art
|
|
HDV genotype sensitivity
|
all relevant genotypes/subtypes, preferably from international reference materials
potential substitutes for rare HDV genotypes (to be quantified by appropriate methods): plasmids; synthetic RNA
|
Quantitative NAT: dilution series for demonstration of quantification efficiencies
|
According to the state of the art
|
|
Diagnostic specificity
|
Blood donor specimens
|
Qualitative NAT: ≥100
Quantitative NAT: ≥100
|
According to the state of the art
|
|
Cross-reactivity
|
Potentially cross-reacting specimens
|
|
According to the state of the art
|
|
Carry-over
|
High HDV RNA positive;
HDV RNA negative
|
At least five runs with alternating high-positive and negative specimens shall be performed during robustness studies. The high positive samples shall comprise samples with naturally occurring high virus titres.
|
According to the state of the art
|
|
Whole system failure rate
|
HDV RNA low positive
|
≥100 HDV RNA low-positive specimens shall be tested. These specimens shall contain a virus concentration equivalent to three times the 95 % positive cut-off virus concentration.
|
≥99% positive
|
ANNEX VIII
COMMON SPECIFICATIONS FOR DEVICES INTENDED FOR DETECTION OF MARKERS OF VARIANT CREUTZFELDT-JACOB (vCJD) DISEASE
Scope
This Annex applies to devices intended for detection of markers of variant Creutzfeldt-Jakob disease (vCJD).
Table 1 applies to devices intended for detection of markers of vCJD.
Table 1. Devices for detection of markers of vCJD
|
Performance characteristic
|
Material
|
Number of specimens
|
Acceptance criteria
|
|
Analytical sensitivity
|
vCJD brain spikes in human plasma (WHO reference number NHBY0/0003)
|
≥24 replicates of each of three dilutions of the material WHO number NHBY0/0003 (1×104, 1×105, 1×106)
|
23 of the 24 replicates detected at 1×104
|
|
|
vCJD spleen spikes in human plasma (10% spleen homogenate — NIBSC reference number NHSY0/0009)
|
≥24 replicates of each of three dilutions of the material NIBSC number NHSY0/0009 (1×10, 1×102 , 1×103 )
|
23 of the 24 replicates detected at 1×10
|
|
Diagnostic sensitivity
|
Specimens from appropriate animal models
|
As many specimens as reasonably possible and available, and ≥10 specimens
|
90%
|
|
|
Specimens from humans with known clinical vCJD
|
As many specimens as reasonably possible and available, and ≥10 specimens
|
90%
|
|
|
|
Only in cases where 10 specimens are not available:
— the number of specimens tested shall be between 6 and 9
— all available specimens shall be tested
|
max. one false negative result
|
|
Analytical specificity
|
Potentially cross-reacting blood-specimens
|
≥100
|
|
|
Diagnostic specificity
|
Normal human plasma samples from area of low bovine spongiform encephalopathy (BSE) exposure
|
≥5000
|
≥ 99,5 %
|
ANNEX IX
COMMON SPECIFICATIONS FOR DEVICES INTENDED FOR DETECTION OR QUANTIFICATION OF MARKERS OF CYTOMEGALOVIRUS (CMV) INFECTION
Scope
This Annex applies to devices intended for detection or quantification of markers of cytomegalovirus (CMV) infection.
Table 1 applies to first-line assays for total antibodies against CMV (total anti-CMV) and IgG antibodies against CMV (anti-CMV IgG).
Table 2 applies to qualitative and quantitative NAT devices for CMV DNA.
Table 1. First-line assays: total anti-CMV and anti-CMV IgG
|
Performance characteristic
|
Specimens
|
Total anti-CMV, anti-CMV IgG
|
Acceptance criteria
|
|
Diagnostic sensitivity
|
Positive specimens
|
≥400
including specimens from recent and past CMV infection,
low and high positive titre samples
|
≥ 99% sensitivity for confirmable past infection
XXIII
;
overall sensitivity including recent infection
XXIV
shall be comparable to other CE-marked tests
|
|
|
Seroconversion panels
|
To be tested when available
|
Diagnostic sensitivity during seroconversion shall represent the state of the art
|
|
Analytical sensitivity
|
Standards
|
WHO international standard anti-CMV IgG (PEI-code 136616/17)
In case of titre determinations and quantitative statements
|
|
|
Diagnostic specificity
|
Negative specimens
|
≥400
XXV
CMV negatives from unselected donors, as compared to another CMV test.
|
≥ 99%
|
|
|
Hospitalised patients
XXVI
|
≥200
|
Potential limitations for specificity, if any, shall be identified
|
|
Cross-reactivity
|
Potentially cross-reacting
XXVII
specimens
|
≥100 in total
(e.g. RF+, related viruses or other infectious agents, pregnant women, etc.)
|
|
Table 2. Qualitative and quantitative NAT devices for CMV DNA
1.For target sequence amplification devices, a functionality control for each specimen (internal control) shall reflect the state of the art. This control shall as far as possible be used throughout the whole process, i.e. extraction, amplification/hybridisation, detection.
2.Genotype and/or subtype detection shall be demonstrated by appropriate primer or probe design validation and shall also be validated by testing characterised genotyped samples.
3.Potential cross-reactivity of non-target nucleic acid sequences shall be analysed by appropriate primer or probe design validation and shall also be validated by testing selected samples.
4.Results of quantitative NAT devices shall be traceable to international standards or calibrated reference materials, if available, and be expressed in international units utilised in the specific field of application.
|
Performance characteristic
|
Specimens
|
Specimen numbers, features, use
|
Acceptance criteria
|
|
Analytical sensitivity
|
WHO 1st International Standard Human CMV DNA (09/162; 5,000,000 IU/vial) (or calibrated reference materials)
|
NAT sensitivity and NAT LOD shall be validated by dilution series of reference materials, testing of replicates (minimal 24) at different analyte concentrations, including those with transition from positive to negative results in the respective NAT assay.
LOD shall be expressed as 95% positive cut-off value (IU/ml) after statistical analysis (e.g. Probit).
XXVIII
quantitative NAT: definition of lower, upper quantification limit, precision, accuracy,
‘linear’ measuring range, ‘dynamic range’.
Reproducibility at different concentration levels
|
According to the state of the art
|
|
Diagnostic sensitivity
CMV Strain sensitivity
|
Patient samples determined as CMV DNA positive by comparator device
Dilution series of CMV positive cell cultures may serve as potential substitutes
|
Qualitative NAT: ≥100
Quantitative NAT: ≥100
dilution series for demonstration of quantification efficiencies
|
According to the state of the art
|
|
Diagnostic specificity
|
Blood donor specimens
|
Qualitative NAT: ≥500
Quantitative NAT: ≥100
|
According to the state of the art
|
|
Cross-reactivity
|
Potentially cross-reacting specimens
|
≥20
Including human samples positive for related human herpesviruses, e.g. EBV, HHV6, VZV
Herpesvirus positive cell cultures may serve as potential substitutes
|
According to the state of the art
|
|
Carry-over
|
High CMV DNA positive;
CMV DNA negative
|
At least five runs with alternating high-positive and negative specimens shall be performed during robustness studies. The virus titres of the high positive samples shall be representative of high virus titres occurring naturally.
|
According to the state of the art
|
|
Whole system failure rate
|
CMV DNA low positive
|
≥100 CMV DNA low-positive specimens shall be tested. These specimens shall contain a virus concentration equivalent to three times the 95 % positive cut-off virus concentration.
|
≥99% positive
|
ANNEX X
COMMON SPECIFICATIONS FOR DEVICES INTENDED FOR DETECTION OR QUANTIFICATION OF MARKERS OF EPSTEIN-BARR VIRUS (EBV) INFECTION
Scope
This Annex applies to devices intended for detection or quantification of markers of Epstein-Barr virus (EBV) infection.
Table 1 applies to first-line assays for IgG antibodies against viral capsid antigen of EBV (anti-EBV VCA IgG).
Table 2 applies to qualitative and quantitative NAT devices for EBV DNA.
Table 1: First-line assays: anti-EBV VCA IgG
|
Performance characteristic
|
Specimens
|
Anti-EBV VCA IgG
|
Acceptance criteria
|
|
Diagnostic sensitivity
|
Positive specimens
|
≥400
including specimens from recent and past EBV infection,
low and high positive titre samples
|
≥ 99% for confirmable past infection
XXIX
; overall sensitivity including recent infection
XXX
shall be equivalent to other CE-marked tests
|
|
|
Seroconversion panels
|
To be tested when available
|
diagnostic sensitivity during seroconversion shall represent the state of the art
|
|
Analytical sensitivity
|
Standards
|
International reference reagents, when available
|
|
|
Diagnostic specificity
|
Negative specimens
|
≥ 200
XXXI
EBV negatives from unselected donors as compared to another EBV test.
|
≥ 99%
|
|
|
Hospitalised patients
XXXII
|
≥200
|
Potential limitations for specificity, if any, shall be identified
|
|
Cross-reactivity
|
Potentially cross-reacting specimens
|
≥100 in total
(e.g. RF+, related viruses or other infectious agents, pregnant women, etc.)
|
|
Table 2. Qualitative and quantitative NAT devices for EBV DNA
1.For target sequence amplification devices, a functionality control for each specimen (internal control) shall reflect the state of the art. This control shall as far as possible be used throughout the whole process, i.e. extraction, amplification/hybridisation, detection.
2.Genotype and/or subtype detection shall be demonstrated by appropriate primer or probe design validation and shall also be validated by testing characterised genotyped samples.
3.Potential cross-reactivity of non-target nucleic acid sequences shall be analysed by appropriate primer or probe design validation and shall also be validated by testing selected samples.
4.Results of quantitative NAT devices shall be traceable to international standards or calibrated reference materials, if available, and be expressed in international units utilised in the specific field of application.
|
Performance characteristic
|
Specimens
|
Specimen numbers, features, use
|
Acceptance criteria
|
|
Analytical sensitivity
|
WHO 1st International Standard Human EBV DNA (09/260; 5,000,000 IU/vial) (or calibrated reference materials)
|
NAT sensitivity and NAT LOD shall be validated by dilution series of reference materials, testing of replicates (minimal 24) at different analyte concentrations, including those with transition from positive to negative results in the respective NAT assay.
LOD shall be expressed as 95% positive cut-off value (IU/ml) after statistical analysis (e.g. Probit).
XXXIII
quantitative NAT: definition of lower, upper quantification limit, precision, accuracy,
‘linear’ measuring range, ‘dynamic range’.
Reproducibility at different concentration levels
|
According to the state of the art
|
|
Diagnostic sensitivity
EBV strain sensitivity
|
Patient samples determined as EBV DNA positive by comparator device
Dilution series of EBV positive cell cultures may serve as potential substitutes
|
Qualitative NAT: ≥100
Quantitative NAT: ≥100
dilution series for demonstration of quantification efficiencies
|
|
|
Diagnostic specificity
|
Negative specimens
|
Qualitative NAT: ≥500
Quantitative NAT: ≥100
|
According to the state of the art
|
|
Cross-reactivity
|
Potentially cross-reacting specimens
|
≥20
Including human samples positive for related human herpesviruses, e.g. CMV, HHV6, VZV
Herpesvirus positive cell cultures may serve as potential substitutes
|
According to the state of the art
|
|
Carry-over
|
High EBV DNA positive;
EBV DNA negative
|
At least five runs with alternating high-positive and negative specimens shall be performed during robustness studies. The virus titres of the high positive samples shall be representative of high virus titres occurring naturally.
|
According to the state of the art
|
|
Whole system failure rate
|
EBV DNA low positive
|
≥100 EBV DNA low-positive specimens shall be tested. These specimens shall contain a virus concentration equivalent to three times the 95 % positive cut-off virus concentration.
|
≥99% positive
|
ANNEX XI
COMMON SPECIFICATIONS FOR DEVICES INTENDED FOR DETECTION OF MARKERS OF TREPONEMA PALLIDUM INFECTION
Scope
This Annex applies to devices intended for detection of Treponema pallidum (T. pallidum).
Table 1 applies to first-line assays for antibodies against T. pallidum (anti-T.pallidum).
Table 2 applies to confirmatory and supplemental anti-T.pallidum assays.
Table 1. First-line assays: anti-T.pallidum
|
Performance characteristic
|
Specimens
|
Specimen number, features, use
|
Acceptance criteria
|
|
Diagnostic
sensitivity
|
Positive specimens
|
≥200 positive samples in total,
at different stages of the infection if available,
including high positive and weakly positive samples,
identified as positive by at least two different serological tests (one of which is an enzyme immunoassay) for different antibodies to T.pallidum
|
≥99.5% overall sensitivity
|
|
|
Seroconversion panels
|
At least 1 seroconversion panel, ≥1 if possible, including individual samples from the early infection phase
|
Diagnostic sensitivity during seroconversion shall represent the state of the art.
|
|
Analytical
sensitivity
|
Standards
|
WHO international standards
NIBSC code 05/132, when available
|
|
|
Diagnostic specificity
|
Unselected blood donors (including first-time donors)
XXXIV
|
≥5000
|
≥99,5%
|
|
|
Hospitalised patients
|
≥200
|
Potential limitations for specificity, if any, shall be identified
|
|
Cross-reactivity
|
Potentially cross-reacting blood- specimens
|
≥100 in total
including the following specimens: positive for Borrelia burgdorferi sensu lato confirmed by IgG immunoblot; anti-HIV positive; RF+; other related microbial/infectious agents; systemic lupus erythematosus (SLE) patients; antiphospholipid antibody positive; pregnant women etc.
|
|
Table 2. Confirmatory and supplemental assays: anti-T.pallidum
|
Performance characteristic
|
Specimens
|
Specimen number, features, use
|
Acceptance criteria
|
|
Diagnostic
sensitivity
|
Positive specimens
|
≥300 positive samples at different stages of the infection (early primary syphilis, secondary stage, and during late syphilis ) including high positive samples, 50 weakly positive samples,
by at least two different serological tests (one of which is an enzyme immunoassay) for different antibodies to T.pallidum
|
99% identification as “confirmed positive” or “indeterminate”
|
|
|
Seroconversion panels
|
At least 1 seroconversion panel , ≥1 if possible, including individual samples from the early infection phase
|
diagnostic sensitivity during seroconversion shall represent the state of the art
|
|
Analytical
sensitivity
|
Standards
|
WHO international standards
NIBSC code 05/132
|
|
|
Diagnostic
specificity
|
Blood donors
|
≥200
|
≥ 99%;
|
|
|
Clinical samples
|
≥200
|
Potential limitations for specificity, if any, shall be identified
|
|
Cross-reactivity
|
Potentially cross-reacting specimens
|
≥50 in total, including samples from pregnant women and samples with indeterminate results in other confirmatory assays.
|
|
ANNEX XII
COMMON SPECIFICATIONS FOR DEVICES INTENDED FOR DETECTION OR QUANTIFICATION OF MARKERS OF TRYPANOSOMA CRUZI INFECTION
Scope
This Annex applies for devices intended for detection or quantification of markers of Trypanosoma cruzi (T. cruzi) infection.
Table 1 applies to first-line assays for antibodies against T. cruzi (anti-T. cruzi).
Table 2 applies to confirmatory and supplemental anti-T. cruzi assays.
Table 3 applies to qualitative and quantitative NAT devices for T. cruzi DNA.
Table 1. First-line assays: anti-T. cruzi
|
Performance characteristic
|
Specimens
|
Specimen number, features, use
|
Acceptance criteria
|
|
Diagnostic
sensitivity
|
Positive specimens
|
≥400 positive samples, including highly positive confirmed by at least two different serological tests for different antibodies to T.cruzi.
Of those 400, ≥25 parasite positive samples, which have been confirmed by direct detection.
|
99.5% overall sensitivity
|
|
|
Seroconversion panels
|
To be defined when available
|
Diagnostic sensitivity during seroconversion shall represent the state of the art
|
|
Analytical
sensitivity
|
Standards
|
WHO international standards
NIBSC code: 09/186
NIBSC code: 09/188
|
|
|
Diagnostic specificity
|
Unselected donors (including first-time donors)
XXXV
|
≥5000
|
≥99,5%
|
|
|
Hospitalised patients
|
≥200
|
Potential limitations for specificity, if any, shall be identified
|
|
Cross-reactivity
|
Potentially cross-reacting specimens
|
≥100 in total
including the following specimens: positive for anti-Toxoplasma gondii; at least 5 specimens positive for anti-Leishmania; RF+; related microbial agents or other infectious agents; SLE patients; antiphospholipid antibody positive patients; pregnant women, etc.
|
|
Table 2. Confirmatory and supplemental assays: anti-T. cruzi
|
Performance characteristic
|
Specimens
|
Specimen number, features, use
|
Acceptance criteria
|
|
Diagnostic
sensitivity
|
Positive specimens
|
≥300 positive specimens, including highly positive confirmed by at least two different serological tests for different antibodies to T.cruzi.
Of those 300, ≥25 parasite positive samples, which have been confirmed by direct detection.
|
≥99% identification as “confirmed positive” or “indeterminate”
|
|
|
Seroconversion panels
|
As available
|
Diagnostic sensitivity during seroconversion shall represent the state of the art, if applicable
|
|
Analytical
sensitivity
|
Standards
|
WHO international standards
NIBSC code: 09/186
NIBSC code: 09/188
|
|
|
Diagnostic
specificity
|
Negative specimens
|
≥200
|
≥99%
|
|
|
Clinical samples
|
≥200
|
Potential limitations for specificity, if any, shall be identified
|
|
Cross-reactivity
|
Potentially cross-reacting specimens
|
≥50, including samples from pregnant women and samples with indeterminate results in other confirmatory assays
|
|
Table 3: NAT devices for T. cruzi DNA
1.For target sequence amplification devices, a functionality control for each specimen (internal control) shall reflect the state of the art. This control shall as far as possible be used throughout the whole process, i.e. extraction, amplification/hybridisation, detection.
2.Genotype and/or subtype detection shall be demonstrated by appropriate primer or probe design validation and shall also be validated by testing characterised genotyped samples.
3.Potential cross-reactivity of non-target nucleic acid sequences shall be analysed by appropriate primer or probe design validation and shall also be validated by testing selected samples.
4.Results of quantitative NAT devices shall be traceable to international standards or calibrated reference materials, if available, and be expressed in international units utilised in the specific field of application.
|
Performance characteristic
|
Specimens
|
Specimen number, features, use
|
Acceptance criteria
|
|
Analytical sensitivity
|
Characterized in-house reference preparation (as long as international reference materials are not available)
|
NAT sensitivity and NAT LOD shall be validated by dilution series of reference materials, testing of replicates (minimal 24) at different analyte concentrations, including those with transition from positive to negative results in the respective NAT assay.
LOD shall be expressed as 95% positive cut-off value (IU/ml) after statistical analysis (e.g. Probit).
|
According to the state of the art
|
|
Diagnostic sensitivity: different T.cruzi strains / isolates
|
Patient samples from different regions determined as T.cruzi DNA positive by comparator device; sequence variants
|
≥100
Dilution series of T.cruzi positive cell cultures (isolates) or T.cruzi positive materials from animal models may serve as potential substitutes
|
According to the state of the art
|
|
Diagnostic specificity
|
Negative specimens
|
≥100
|
According to the state of the art
|
|
Cross-reactivity
|
Potentially cross-reacting specimens
|
≥10 human samples positive for other parasites, e.g. Plasmodium species, Trypanosoma brucei. Positive cell cultures may serve as potential substitutes
|
According to the state of the art
|
|
Carry-over
|
|
At least five runs with alternating high-positive and negative specimens shall be performed during robustness studies. The T.cruzi titres of the high positive samples shall be representative of high T.cruzi titres occurring naturally.
|
According to the state of the art
|
|
Whole system failure rate
|
|
≥100 T.cruzi DNA low-positive specimens shall be tested. These specimens shall contain a T.cruzi concentration equivalent to three times the 95 % positive cut-off T.cruzi concentration.
|
≥99% positive
|
ANNEX XIII
COMMON SPECIFICATIONS FOR DEVICES INTENDED FOR DETECTION OR QUANTIFICATION OF MARKERS OF SEVERE ACUTE RESPIRATORY SYNDROME CORONAVIRUS 2 INFECTION
Scope
This Annex applies to devices intended for detection or quantification of markers of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2)infection.
Table 1 applies to the following first-line assays (including rapid tests) for antibodies against SARS-CoV-2 (anti-SARS-CoV-2): total antibody, IgG-only, IgG combined with IgM and/or IgA.
Table 2 applies to first-line assays (including rapid tests) for detection of anti-SARS-CoV-2 IgM and/or IgA.
Table 3 applies to confirmatory or supplemental assays for anti-SARS-CoV-2.
Table 4 applies to antigen SARS-CoV-2 tests, including rapid antigen tests.
Table 5 applies to NAT assays for SARS-CoV-2 RNA.
Table 6 applies to SARS-CoV-2 antigen self-tests which have already undergone a performance evaluation for professional use.
Table 7 applies to SARS-CoV-2 antibody self-tests which have already undergone a performance evaluation for professional use.
Table 1: First-line assays (including rapid tests) for anti-SARS-CoV-2: total antibody, IgG-only, IgG combined
XXXVI
with IgM and/or IgA
|
Performance characteristic
|
Specimen
|
Anti-SARS-CoV-2 total antibody, IgG, IgG combined
|
Acceptance criteria
|
|
Diagnostic sensitivity
|
Positive specimens
|
≥400
including samples from early infection and post seroconversion
XXXVII
(within the first 21 days and after 21 days following the onset of symptoms);
including samples from asymptomatic or subclinical and mildly symptomatic (outpatient treatment) individuals;
including samples with low and high titers;
including samples from vaccinated individuals where appropriate
XXXVIII
;
consideration of genetic variants
|
≥90% sensitivity
XXXIX
for samples taken >21 days after onset of symptoms
XL
;
overall sensitivity including the early infection phase shall be comparable to other CE marked
XLI
tests
|
|
|
Seroconversion panels
|
As far as available
|
Seroconversion sensitivity comparable to other CE-marked tests
|
|
Analytical sensitivity
|
Reference preparations
|
WHO International Standard (IS) for anti- SARS-CoV-2 (NIBSC code 20/136);
WHO International Reference Panel (RP) for anti-SARS-CoV-2 antibodies (NIBSC codes 20/140, 20/142, 20/144, 20/148, 20/150)
|
IS: for titre determinations / quantitative
XLII
result output;
RP: all antibody assays
|
|
Specificity
|
Negative specimens
XLIII
|
≥400
samples from non-infected and non-vaccinated individuals
XLIV
|
>99% specificity
XLV
|
|
|
|
≥200
hospitalised patients (without SARS-CoV-2 infection)
|
Potential limitations for specificity shall be determined
|
|
|
|
≥100 in total
potentially interfering (e.g. rheumatoid factor, pregnant women, etc.) and cross-reacting blood specimens: including antibodies against endemic human coronaviruses 229E, OC43, NL63, HKU1 and other pathogens of respiratory diseases such as influenza A, B, RSV etc.
|
|
Table 2: First-line assays (including rapid tests) for anti-SARS-CoV-2: IgM and/or IgA detection
|
Performance characteristic
|
Specimen
|
Anti-SARS-CoV-2 IgM and/or IgA
|
Acceptance criteria
|
|
Diagnostic sensitivity
|
Positive specimens
|
≥200
XLVI
samples
XLVII
with a significant proportion from the early phase of the infection (within 21 days after onset of symptoms) compared to samples past seroconversion (>21 days after onset of symptoms);
including samples from asymptomatic, subclinical, mildly symptomatic (outpatient treatment) individuals;
including freshly
XLVIII
vaccinated individuals if appropriate;
consideration of genetic variants
|
≥80% sensitivity
XLIX
for samples taken during the first 21 days after symptom onset
L
;
overall sensitivity shall be comparable to other CE marked
LI
tests of the same type (i.e. IgM and/or IgA)
|
|
|
Seroconversion panels
|
As far as available
|
Seroconversion sensitivity comparable to other CE-marked tests
|
|
Analytical sensitivity
|
Standards
|
N/A
|
N/A
|
|
Specificity
|
Negative specimens
LII
|
≥200
samples from non-infected and non-vaccinated individuals
LIII
|
≥98% specificity
LIV
|
|
|
|
≥100
from hospitalised patients (without SARS-CoV-2 infection)
|
Potential limitations for specificity shall be determined
|
|
|
|
≥100 in total
potentially interfering (e.g. rheumatoid factor, pregnant women, etc.) and cross-reacting blood specimens; antibodies against endemic human coronaviruses 229E, OC43, NL63, HKU1 and other pathogens of respiratory diseases such as influenza A, B, RSV etc.
|
|
Table 3: Confirmatory or supplemental
LV
assays for anti-SARS-CoV-2
|
Performance characteristic
|
Specimen
|
Anti-SARS-CoV-2
|
Acceptance criteria
|
|
Diagnostic sensitivity
|
Positive specimens
|
≥200
including samples pre and post seroconversion (within the first 21 days and after 21 days following the onset of symptoms)
|
Correct determination as “positive” (or “indeterminate”)
|
|
|
Seroconversion panels/
low titre panels
|
as far as available
|
|
|
Analytical sensitivity
|
Standards
|
N/A
|
N/A
|
|
Diagnostic specificity
|
Negative specimens
LVI
|
≥200 from non-infected / non-vaccinated population
|
No false positive results;
correct determination as “negative” (or “indeterminate”)
|
|
|
|
≥200 from hospitalised patients (without SARS-CoV-2 infection)
≥50 potentially interfering and cross-reacting samples in total: antibodies against endemic human coronaviruses 229E, OC43, NL63, HKU1 and other pathogens of respiratory diseases such as influenza A, B, RSV etc.;
including samples with indeterminate or false positive results in other anti-SARS-CoV-2 assays
|
|
Table 4: Antigen assays (including rapid tests): SARS-CoV-2
|
Performance characteristic
|
Specimen
|
SARS-CoV-2 antigen
|
Acceptance criteria
|
|
Diagnostic sensitivity
|
Positive specimens
|
≥100
LVII
NAT positive samples
LVIII
from early infection within the first 7 days after symptom onset
LIX
;
samples shall represent naturally occurring viral loads
LX
;
consideration of genetic variants
LXI
;
consideration of variations in specimen collection and/or specimen handling
LXII
|
Detection of >80% (rapid tests);
detection of >85% (lab-based assays
LXIII
);
relative to SARS-CoV-2-NAT
LXIV
,
LXV
|
|
Analytical sensitivity
|
Standards
|
As soon as available
|
Establishment of a LOD
LXVI
|
|
Diagnostic specificity
|
Negative specimens
|
≥300
from non-infected individuals
|
Specificity >98% (rapid tests)
Specificity >99% (lab-based assays7)
|
|
|
|
≥100 from hospitalised patients
≥50 potentially interfering and cross-reactive samples in total: including virus-positive samples of endemic human coronaviruses 229E, OC43, NL63, HKU1; influenza A, B, RSV, and other pathogens of respiratory diseases, eligible for differential diagnosis; including bacteria
LXVII
present in the sampling area
|
Potential limitations for specificity shall be determined
|
Table 5: NAT assays for SARS-CoV-2 RNA
|
Performance characteristic
|
Specimen
|
SARS-CoV-2 RNA qualitative
|
SARS-CoV-2 RNA quantitative
|
|
Sensitivity
|
|
Analytical sensitivity: LOD
|
WHO 1st International Standard SARS-CoV-2 RNA (NIBSC code 20/146; 7.70 Log10 IU/mL)
Secondary standards calibrated against WHO IS
|
According to Ph. Eur. NAT validation guideline:
several dilution series into borderline concentration; statistical analysis (e.g. Probit analysis) on the basis of at least 24 replicates; calculation of 95 % cut-off value
|
According to Ph. Eur. NAT validation guideline:
several dilution series of calibrated reference preparations into borderline concentration; statistical analysis (e.g. Probit analysis) on the basis of at least 24 replicates; calculation of 95 % cut-off value as LOD
|
|
Quantification limit; quantification features
|
WHO 1st International Standard SARS-CoV-2 RNA (NIBSC code 20/146; 7.70 Log10 IU/mL)
Secondary standards calibrated against WHO IS
|
|
Dilutions (half-log10 or less) of calibrated reference preparations; determination of lower, upper quantification limit, LOD, precision, accuracy, “linear” measuring range, “dynamic range”. Synthetic target may be used as secondary standard to achieve higher concentration levels. Reproducibility at different concentration levels to be shown
|
|
Diagnostic sensitivity: different SARS-CoV-2 RNA strains
|
Patient samples determined as SARS-CoV-2 RNA positive by comparator device from different regions and outbreak clusters; sequence variants
Dilution series of SARS-CoV-2 positive cell cultures (isolates) may serve as potential substitutes
|
≥100
LXVIII
|
|
|
Quantification efficiency
|
SARS-CoV-2 RNA positive patient samples from different regions and outbreak clusters; sequence variants
with quantitative values obtained by comparator device
Dilution series of SARS-CoV-2 RNA positive cell cultures may serve as potential substitutes
|
|
≥100
|
|
Inclusivity
|
In silico analysis
LXIX
;
at least two independent target gene regions in one test run (dual-target design)
|
Evidence of suitable assay design: primer/probe sequence alignments with published SARS-CoV-2 sequences
|
Evidence of suitable assay design: primer/probe sequence alignments with published SARS-CoV-2 sequences
|
|
Specificity
|
|
Diagnostic specificity
|
SARS-CoV-2 RNA negative human specimens
|
≥500
|
≥100
|
|
In silico analysis2
|
|
Evidence of suitable assay design evidence (sequence alignments); regular check of primer/probe sequences against sequence data bank entries
|
Evidence of suitable assay design evidence (sequence alignments); regular check of primer/probe sequences against sequence data bank entries
|
|
Potential cross reaction
|
samples positive (various concentrations) for related human coronaviruses 229E, HKU1, OC43, NL63, MERS coronavirus; SARS CoV-1 if available; Influenza virus A, B; RSV; Legionella pneumophila;
positive cell cultures may serve as potential substitutes
|
≥20 in total
|
≥20
|
|
Robustness
|
|
Cross contamination
|
|
At least 5 runs using alternating high positive and negative samples. The virus titres of the high positive samples shall be representative of high virus titres occurring naturally.
|
At least 5 runs using alternating high positive (known to occur naturally) and negative samples
|
|
Inhibition
|
|
Internal control preferably to go through the whole NAT procedure
|
Internal control preferably to go through the whole NAT procedure
|
|
Whole system failure rate leading to false negative results: 99/100 assays positive
|
|
≥100 samples virus-spiked with 3 × the 95 % positive cut-off concentration (3 x LOD)
|
≥100 samples virus-spiked with 3 × the 95 % positive cut-off concentration (3 x LOD)
|
Table 6: Additional requirements for SARS-CoV-2 antigen self-tests
LXX
|
Performance characteristic
|
Specimens
LXXI
|
Number of lay users
|
Criterion
|
|
Result interpretation
|
Interpretation of contrived tests
LXXII
by lay users reflecting a range of results:
·non-reactive
·reactive
·weak reactive
LXXIII
·invalid
|
≥100
|
Reading and interpretation of the contrived test results by 100 lay people;
each lay person shall be subjected to read the specified range of result reactivity levels;
determination of concordance of lay reading of the same tests by professional readers
|
|
Diagnostic sensitivity
|
Lay users that are known antigen positive
LXXIV
,
LXXV
|
≥30
|
In comparison to the true infectious status, i.e. by RT-PCR;
concordance of results with the professional test
|
|
Diagnostic specificity
|
Lay users that do not know their status5
|
≥60
|
Concordance of results with the professional test
|
Table 7: Additional requirements for SARS-CoV-2 antibody self-tests
LXXVI
|
Performance characteristic
|
Specimens
LXXVII
|
Number of lay users
|
Criterion
|
|
Result interpretation
|
Interpretation of contrived tests
LXXVIII
by lay users reflecting a range of results:
·non-reactive
·reactive
·weak reactive
LXXIX
·invalid
|
≥100
|
Reading and interpretation of the contrived test results by 100 lay people;
each lay person shall be subjected to read the specified range of result reactivity levels;
determination of concordance of lay reading of the same tests by professional readers
|
|
Diagnostic sensitivity
|
Lay users that are known antibody positive
LXXX
|
≥100
|
With previous history of initial PCR confirmed infection for SARS-CoV-2;
in comparison to a previous confirmed antibody result;
concordance of results with the professional test
|
|
Diagnostic specificity
|
Lay users that do not know their status5
|
≥100
|
Concordance of results with the professional test
|
-
(I)
Only where reactivity against these antigens is claimed.
-
(II)
Blood donor populations shall be investigated from at least two blood donation centres and consist of consecutive blood donations, which have not been selected to exclude first-time donors.
-
(III)
Reference: European Pharmacopeia 9.0, 2.6.21 Nucleic acid amplification techniques, Validation.
-
(IV)
For each body fluid claimed for use with the device, e.g. whole blood, urine, saliva, etc. The sensitivity and specificity of the device for self-testing in the hands of lay users shall be defined against the confirmed patient infectious status.
-
(V)
The result interpretation study shall include reading and interpretation of the contrived test results by 100 lay people with each lay person subjected to read one or more contrived tests from the specified range of result reactivity levels. The manufacturer shall determine the concordance between lay user reading and professional user reading.
-
(VI)
Tests shall be performed prior to the result interpretation study using whenever possible the specimen type intended by the manufacturer.
-
(VII)
A higher proportion of the specimens shall be in the weak-positive range close to the cut-off or LOD of the test.
-
(VIII)
Blood donor populations shall be investigated from at least two blood donation centres and consist of consecutive blood donations, which have not been selected to exclude first-time donors.
-
(IX)
Reference: European Pharmacopeia 9.0, 2.6.21 Nucleic acid amplification techniques, Validation.
-
(X)
Blood donor populations shall be investigated from at least two blood donation centres and consist of consecutive blood donations, which have not been selected to exclude first-time donors.
-
(XI)
Reference: European Pharmacopeia 9.0, 2.6.21 Nucleic acid amplification techniques, Validation.
-
(XII)
For each body fluid claimed for use with the device, e.g. whole blood, urine, saliva, etc. The sensitivity and specificity of the device for self-testing in the hands of lay users shall be defined against the confirmed patient infectious status.
-
(XIII)
The result interpretation study shall include reading and interpretation of the contrived test results by 100 lay people with each lay person subjected to read one or more contrived tests from the specified range of result reactivity levels. The manufacturer shall determine the concordance between lay user reading and professional user reading.
-
(XIV)
Tests shall be performed prior to the result interpretation study using whenever possible the specimen type intended by the manufacturer.
-
(XV)
A higher proportion of the specimens shall be in the weak-positive range close to the cut-off or LOD of the test.
-
(XVI)
Blood donor populations shall be investigated from at least two blood donation centres and consist of consecutive blood donations, which have not been selected to exclude first-time donor.
-
(XVII)
Reference: European Pharmacopeia 9.0, 2.6.21 Nucleic acid amplification techniques, Validation.
-
(XVIII)
For each body fluid claimed for use with the device, e.g. whole blood, urine, saliva, etc. The sensitivity and specificity of the device for self-testing in the hands of lay users shall be defined against the confirmed patient infectious status.
-
(XIX)
The result interpretation study shall include reading and interpretation of the contrived test results by 100 lay people with each lay person subjected to read one or more contrived tests from the specified range of result reactivity levels. The manufacturer shall determine the concordance between lay user reading and professional user reading.
-
(XX)
Tests shall be performed prior to the result interpretation study using whenever possible the specimen type intended by the manufacturer.
-
(XXI)
A higher proportion of the specimens shall be in the weak-positive range close to the cut-off or LOD of the test.
-
(XXII)
Reference: European Pharmacopeia 9.0, 2.6.21 Nucleic acid amplification techniques, Validation.
-
(XXIII)
Including testing of other CMV parameters (e.g. CMV-IgM, avidity, immunoblot), or previous / follow-up samples for true sample status.
-
(XXIV)
Supplementary testing to confirm recent CMV infection (primary or re-infection): e.g. CMV-IgM, IgG-avidity, immunoblot analysis.
-
(XXV)
Corresponding to an initial number of 1000 donors at an assumed CMV prevalence of 60 %.
-
(XXVI)
Including pre-transplant recipients.
-
(XXVII)
Including related ß-herpes viruses (HHV-6, HHV-7).
-
(XXVIII)
Reference: European Pharmacopeia 9.0, 2.6.21 Nucleic acid amplification techniques, Validation.
-
(XXIX)
Including other EBV markers and parameters (e.g. VCA-IgM, EBNA-1 IgG, immunoblot) to assess the true specimen status.
-
(XXX)
Supplementary testing to confirm recent EBV infection: e.g. VCA-IgM, IgG-avidity, immunoblot analysis.
-
(XXXI)
At an assumed EBV prevalence of 80 % corresponding to an initial number of 1000 donors.
-
(XXXII)
Including pre-transplant recipients.
-
(XXXIII)
Reference: European Pharmacopeia 9.0, 2.6.21 Nucleic acid amplification techniques, Validation.
-
(XXXIV)
Blood donor populations shall be investigated from at least two blood donation centres and consist of consecutive blood donations, which have not been selected to exclude first-time donors.
-
(XXXV)
Blood donor populations shall be investigated from at least two blood donation centres and consist of consecutive blood donations, which have not been selected to exclude first-time donors.
-
(XXXVI)
Performance claim of the combined overall result; for devices with separate claims for IgM and/or IgA, see table 2.
-
(XXXVII)
Details on the time interval between sampling and onset of symptoms (or time of infection, if available) shall be provided.
-
(XXXVIII)
The manufacturer shall provide a justification of the suitability and timing for sensitivity evaluation of the relevant antibodies in vaccinated individuals.
-
(XXXIX)
Based on confirmed positive SARS-CoV-2 NAT result.
-
(XL)
Claims for sensitivity shall be specified in relation to the time between sampling after symptom onset or the initial PCR diagnosis and the test.
-
(XLI)
CE marked under Regulation (EU) 2017/746 as class D, if available.
-
(XLII)
This applies to quantitative assays if they are also first-line assays.
-
(XLIII)
Negative specimens shall be from individuals with no history of SARS-CoV-2 infection (if available pre-pandemic).
-
(XLIV)
Individuals vaccinated with an antigen different from that used in the device may be included, if appropriate.
-
(XLV)
False positive results shall be resolved by re-testing in other SARS-CoV-2 serologic assays, if necessary with different test design and antigen coating compared to the initial test, and/or confirmatory testing.
-
(XLVI)
In case of devices detecting both IgM and IgA, 200 per marker IgM and IgA.
-
(XLVII)
Details on the time interval between sampling and onset of symptoms (or time of infection, if available) shall be provided.
-
(XLVIII)
The manufacturer shall provide a justification of the suitability and timing for sensitivity evaluation of IgM and IgA in vaccinated individuals.
-
(XLIX)
Based on confirmed positive SARS-CoV-2 NAT result.
-
(L)
Claims for sensitivity shall be specified in relation to the time between sampling after symptom onset or the initial PCR diagnosis and the test.
-
(LI)
CE marked under Regulation (EU) 2017/746 as class D, if available.
-
(LII)
Negative specimens shall be from individuals with no history of SARS-CoV-2 infection (if available pre-pandemic).
-
(LIII)
Individuals vaccinated with an antigen different from that used in the device may be included, if appropriate.
-
(LIV)
False positive results shall be resolved by re-testing in other SARS-CoV-2 serologic assays, if necessary with different test design and antigen coating compared to the initial test, and/or confirmatory testing. Clarification of false positive results may additionally include testing for presence of other anti-SARS-CoV-2 antibody types (IgA, IgG, total antibody).
-
(LV)
E.g. immunoblot providing antigens different from those used in the initial antibody test.
-
(LVI)
Negative specimens shall be from individuals with no history of SARS-CoV-2 infection (if available pre-pandemic).
-
(LVII)
If the device is intended to be used for more than one specimen type, 100 samples shall be required for each specimen type. If this is not possible in exceptional circumstances (e.g. if specimen collection is very invasive), the manufacturer shall provide a justification and evidence of matrix equivalence.
-
(LVIII)
Sampling shall be matched for antigen and NAT testing, e.g., two simultaneous samples from each individual or optimally NAT- and antigen testing from the same sample (e.g. from the eluate of one swab); the buffer/transport medium shall be compatible for both NAT and antigen testing; any volume change in the buffer/medium for sample uptake different from that of the proprietary assay, and/or between antigen and NAT test shall be clearly communicated.
-
(LIX)
Or time of infection, if known, taking into account the incubation time.
-
(LX)
I.e., without preselection; the viral loads and their distribution shall be shown, e.g. characterized by Ct-values of RT-PCR; or transformed into viral load per ml or sample, if applicable.
-
(LXI)
Depending on the design of the device and nature of the genetic variant. For the purpose of evaluation, at least 3 samples shall be represented for each genetic variant.
-
(LXII)
Specimen collection and extraction items such as swabs, extraction buffers, etc., shall be part of the evaluation. If proprietary sampling/sample preparation is not included in the test kit, test performance shall be investigated for an applicable range of sampling devices. If the sample is not tested immediately, e.g. after a certain transport time, stability of the antigen shall be investigated.
-
(LXIII)
Other than rapid tests, i.e. formal laboratory-based assays e.g. enzyme immunoassay, automated tests, etc.
-
(LXIV)
The sensitivity of ≥80%, ≥85% respectively, shall be for all specimen types claimed. All claimed specimen types shall be compared with paired NAT results from nasopharyngeal specimens.
-
(LXV)
The relationship between antigen test performance and NAT shall be demonstrated; sensitivity may be shown relating to different viral load ranges and to the threshold of infectivity. The NAT and extraction method used shall be described.
-
(LXVI)
Unless there is an available international standard, analytical sensitivity may be tested by dilution series of in-house virus preparations, comparatively with other antigen tests and NAT; if inactivated virus is used, the effect of inactivation and freeze/thawing on the antigen shall be investigated.
-
(LXVII)
E.g. staphylococci and streptococci expressing protein A or G.
-
(LXVIII)
If the device is intended to be used for more than one specimen type, 100 samples shall be required for each specimen type. If this is not possible in exceptional circumstances (e.g. if specimen collection is very invasive), the manufacturer shall provide a justification and evidence of matrix equivalence.
-
(LXIX)
The manufacturer shall define frequency and document evidence of regular surveillance checks against updated data bank entries in a post-market performance follow-up plan and report.
-
(LXX)
It is assumed that the underlying performance of the self-test has already been previously demonstrated with the evaluation/assessment of a professional test of the same design as the respective self-test under evaluation. In case for the self-use specimens in question there is no corresponding professional test variant, comparison shall be made with the standard specimen type (e.g. nasopharyngeal swabs for antigen test, serum or plasma for antibody test) of the corresponding professional test.
-
(LXXI)
For each self-use specimen type claimed with the device (e.g. nasal, sputum, saliva, whole blood, etc.).
-
(LXXII)
Using whenever possible the original natural matrix of the respective specimen type.
-
(LXXIII)
A higher proportion of the samples shall be in the weak-positive range close to the cut-off or LOD of the test.
-
(LXXIV)
Individuals unaware of the professional diagnostic result prior to self-testing, and performing the entire test procedure from specimen collection and specimen pre-treatment (swab, buffer extraction, etc.) to reading.
-
(LXXV)
Subjects up to about 7 days after symptom onset.
-
(LXXVI)
It is assumed that the underlying performance of the self-test has already been previously demonstrated with the evaluation/assessment of a professional test of the same design as the respective self-test under evaluation. In case for the self-use specimens in question there is no corresponding professional test variant, comparison shall be made with the standard specimen type (e.g. nasopharyngeal swabs for antigen test, serum or plasma for antibody test) of the corresponding professional test.
-
(LXXVII)
For each self-use specimen type claimed with the device (e.g. nasal, sputum, saliva, whole blood, etc.).
-
(LXXVIII)
Using whenever possible the original natural matrix of the respective specimen type.
-
(LXXIX)
A higher proportion of the samples shall be in the weak-positive range close to the cut-off or LOD of the test.
-
(LXXX)
Individuals unaware of the professional diagnostic result prior to self-testing, and performing the entire test procedure from specimen collection and specimen pre-treatment (swab, buffer extraction, etc.) to reading.