Ensuring the safety and performance of medical devices

SUMMARY OF:

Regulation (EU) 2017/745 on medical devices

WHAT IS THE AIM OF THE REGULATION?

  • Regulation (EU) 2017/745 updates the rules on placing, making available and putting into service medical devices1 for human use and their accessories on the European Union (EU) market.
  • It also contains rules on how clinical investigations2 concerning such devices and accessories are carried out in the EU.
  • It aims to improve patient safety by introducing procedures for conformity assessment (to ensure that unsafe or non-compliant devices do not end up on the market) and post-market surveillance.
  • Amending Regulation (EU) 2020/561 was adopted to allow EU Member States and their authorities and institutions to prioritise the fight against the COVID‑19 pandemic. The rules deferred the application of certain rules of Regulation (EU) 2017/745 by one year, in order to ensure the smooth functioning of the EU’s internal market, to maintain a high level of protection of public health and patient safety, to provide legal certainty and to avoid potential market disruption during the pandemic.
  • The regulation was later amended by Regulation (EU) 2023/607, extending certain transitional periods, and by Regulation (EU) 2024/1860, which supported the gradual deployment of the European database on medical devices (Eudamed) and required notification of interruptions or discontinuations in the supply of certain devices.

KEY POINTS

Scope

Besides medical devices, the regulation also covers certain groups of products that do not have an intended medical purpose. These include coloured contact lenses (i.e. lenses that do not correct vision) and liposuction equipment. A list of these products is included in Annex XVI to the regulation.

Classification

Medical devices are classified according to their intended purpose and their inherent risks (classes I, IIa, IIb and III as set out in Annex VIII to the regulation).

Notified bodies

  • The regulation tightens the rules concerning how the independent notified bodies – which assess the conformity of medium- and high-risk medical devices before they are placed on the market – are designated, organised and monitored.
  • These bodies have to meet the same high-quality standards throughout the EU and must have the required competences, resources and staff to successfully perform their conformity assessment tasks.
  • On-site inspections of manufacturers, some of which are unannounced, must be carried out.
  • Assessments of certain high-risk devices (e.g. implants) might also involve EU-level panels of independent experts (expert panels).

Clinical data

  • The regulation specifies what is required in the data collection of clinical investigations on medical devices. These requirements have been, to a large extent, aligned with those applicable to clinical trials on medicinal products. They include rules on informed consent and protecting vulnerable subjects (e.g. people under the age of 18, pregnant women or incapacitated people).
  • Clinical investigations conducted in more than one Member State will be subject to a single coordinated assessment.

Obligations of manufacturers and other economic operators

  • Manufacturers have clearer and more stringent obligations to monitor the quality, performance and safety of devices.
  • They must ensure sufficient financial coverage with respect to their potential liability under Directive 85/374/EEC on product liability (see summary) and the related measures must be proportionate to the risk class, the type of device and the size of the enterprise.
  • Manufacturers have to establish quality management and post-market surveillance systems proportionate to the risk class and the type of device.
  • Under the current rules, in the event of damages due to a defective device, a manufacturer’s authorised representative is jointly and severally liable.
  • Where hazardous substances that are carcinogenic, mutagenic or toxic to reproduction or that can interfere with endocrine systems are present in invasive medical devices3 beyond a certain threshold, manufacturers must justify their presence to the notified body.
  • Specific obligations are also established for related economic operators: authorised representatives, importers, distributors and those dealing with systems and procedure packs.

Traceability

The regulation introduces a system for registering devices and manufacturers, importers and authorised representatives to ensure the traceability of devices throughout the supply chain by means of a unique device identifier. This will ensure that measures can be taken rapidly if problems arise.

Single-use devices

The regulation currently allows these devices to be reprocessed (cleaned, disinfected, tested, restored for technical and functional safety and sterilised) only if permitted under national law and if they meet certain conditions laid down in this regulation. Any individual or legal person who reprocesses a single-use device to make it suitable for further use assumes the obligations of a manufacturer. In certain cases, Member States may allow for exceptions to the general rules where the single-use device is reprocessed and used within a health institution, provided that certain specific requirements laid down in the regulation are fulfilled.

A European Commission legislative proposal of would require manufacturers to explain why a device is labelled as single-use and would introduce new rules on when such devices may be reprocessed.

Incident reporting

In addition to the obligation for manufacturers to report serious incidents and trends in non-serious incidents, the regulation introduces obligations for Member States to encourage and enable healthcare professionals, users and patients to report suspected incidents at the national level using standardised formats.

Market surveillance

The relevant EU authorities are responsible for ensuring that any unsafe or non-compliant device is not placed on the market, or is withdrawn from the market if found to be unsafe after being placed on the market.

European database on medical devices

A centralised system, Eudamed, is being developed and progressively implemented to provide Member States, economic operators, patients, healthcare professionals and the public with information on medical devices available in the EU. The rules on the gradual deployment and use of Eudamed were amended by Regulation (EU) 2024/1860. Several electronic systems within Eudamed were declared functional by Decision (EU) 2025/2371, triggering the transition periods for their mandatory use, with the first modules becoming mandatory from .

Implant card

In the case of implantable devices, manufacturers must provide patients with key information on an implant card delivered with the device. This includes:

  • the identification of the device, its name, serial number, lot number, the unique device identifier and the manufacturer’s details;
  • warnings, precautions or measures to be taken regarding possible interference from external influences, medical examinations or environmental conditions;
  • the expected lifetime of the device and any necessary follow-up.

Implementing acts

The full list of implementing acts to Regulation (EU) 2017/745 can be found on this page.

Repeal

Regulation (EU) 2017/745, as amended by Regulation (EU) 2020/561, repealed and replaced Directives 90/385/EEC and 93/42/EEC as of . It lays down specific transitional rules and some exceptions in Articles 120 and 122, which were subsequently amended by Regulation (EU) 2023/607 to extend the transitional periods for certain devices previously certified under those directives.

FROM WHEN DOES THE REGULATION APPLY?

It entered into force on and, further to amending Regulation (EU) 2020/561, has applied from , one year later than originally intended. However, the dates of application for some of the regulation’s rules vary and are detailed in Articles 120, 122 and 123, as amended.

BACKGROUND

  • This regulation is one of two adopted by the EU to overhaul its laws on medical devices. The second regulation (Regulation (EU) 2017/746, see summary) concerns in vitro diagnostic medical devices.
  • For further information, see:

KEY TERMS

  1. Medical device. A term covering a wide variety of products used for medical purposes. It does not achieve its principal intended action by pharmacological, immunological or metabolic means, but the latter may assist in its function. Examples range from bandages to hip replacements to pacemakers. The complete definition of the term medical device is laid down in Article 2(1) of Regulation (EU) 2017/745.
  2. Clinical investigation. A systematic investigation involving one or more human subjects, undertaken to assess the safety or performance of a device.
  3. Invasive medical devices. A device which, wholly or partially, penetrates the body, either through an orifice or through the surface of the body.

MAIN DOCUMENT

Regulation (EU) 2017/745 of the European Parliament and of the Council of on medical devices, amending Directive 2001/83/EC, Regulation (EC) No 178/2002 and Regulation (EC) No 1223/2009 and repealing Council Directives 90/385/EEC and 93/42/EEC (OJ L 117, , pp. 1–175).

Successive amendments to Regulation (EU) 2017/745 have been incorporated into the original text. This consolidated version is of documentary value only.

last update